Full Field Digital, System, X-Ray, Mammographic

FDA product code MUE · Class II (moderate risk) · 21 CFR 892.1715 · Radiology

56510(k) clearances
6PMA approvals
39recalls

Manufacturers

510(k) clearances

510(k)DeviceApplicantDecision
K254090I-View®MUE · Traditional Hologic, Inc.09/02/2026SE
K253678MAMMOMAT B.brilliant (11372950)MUE · Traditional Siemens Medical Solutions USA, Inc.06/18/2026SE
K254092Bellalun 2D (VDMS-1000S)MUE · Traditional Vieworks Co., Ltd.05/28/2026SE
K260087Senographe Pristina; SenoBright HD; Pristina Serena; Pristina Serena BrightMUE · Traditional GE Medical Systems SCS03/24/2026SE
K243420HestiaMUE · Traditional Genoray Co., Ltd.07/17/2025SE
K2438492430TCA with Xmaru WMUE · Special Rayence Co., Ltd.03/12/2025SE
K241113Vivix-MMUE · Traditional Vieworks Co., Ltd.01/16/2025SE
K233539MAMMOMAT B.brilliantMUE · Traditional Siemens Medical Solutions USA, Inc.03/27/2024SE
K220073RMF-2000MUE · Traditional DRTECH Corporation01/26/2023SE
K211720Planmed Clarity 2D, Planmed Clarify SMUE · Traditional Planmed OY07/18/2022SE
K212873Aspire CristalleMUE · Traditional Fujifilm Corporation05/27/2022SE
K210151Digimamo DMUE · Traditional Vmi Tecnologias Ltda03/01/2022SE
K202822HelianthusMUE · Traditional Metaltronica Spa12/15/2021SE
K211725Senographe PristinaMUE · Traditional GE Healthcare08/06/2021SE
K211215SenoBright HDMUE · Traditional GE Healthcare06/22/2021SE
K2029022430MCA with Xmaru WMUE · Traditional Rayence Co., Ltd.06/21/2021SE
K192317Planmed Clarity 2D and Clarity SMUE · Traditional Planmed OY10/23/2020SE
K200424Giotto Class (Models - 3000X-YY and 4000X-YY)MUE · Special Ims Giotto S.P.A.06/18/2020SE
K193166MAMMOMAT RevelationMUE · Traditional Siemens Medical Solutions USA, Inc.06/12/2020SE
K193334Pristina Serena BrightMUE · Traditional GE Healthcare05/15/2020SE
K190809Sample Imaging for Senographe PristinaMUE · Traditional GE Healthcare07/29/2019SE
K182951Pristina Serena 3DMUE · Traditional GE Healthcare01/18/2019SE
K173576Pristina SerenaMUE · Traditional GE Healthcare05/14/2018SE
K173408Mammomat RevelationMUE · Traditional Siemens Medi Cal Solutions, Inc.03/21/2018SE
K163328Planmed ClarityMUE · Traditional Planmed OY12/28/2017SE
K172404SenoBright HDMUE · Traditional GE Healthcare10/30/2017SE
K173132ASPIRE CristalleMUE · Special Fujifilm Medical Systems U.S.A, Inc.10/27/2017SE
K172027Adani MammoScanMUE · Traditional Adani10/20/2017SE
K163302Senographe PristinaMUE · Traditional GE Healthcare09/01/2017SE
K170930RSM 2430CMUE · Special DRTECH Corporation04/27/2017SE
K162670RSM 1824C with RConsole1MUE · Traditional DRTECH Corporation02/23/2017SE
K163252Mammomat Fusion with Stereotactic BiopsyMUE · Traditional Siemens Medical Solutions USA, Inc.01/10/2017SE
K162268Senographe PristinaMUE · Traditional GE Healthcare11/03/2016SE
K151645MAMMOMAT FusionMUE · Traditional Siemens Medical Solutions USA, Inc.09/14/2015SE
K133972Aspire CristalleMUE · Traditional Fujifilm Medical Systems U.S.A, Inc.03/25/2014SE
K130704Microdose SiMUE · Traditional Philips Digital Mammography Sweden AB12/13/2013SE
K131885Senographe Essential, Premium View I (Pvi)MUE · Traditional GE Medical Systems SCS09/26/2013SE
K123520Mammomat Inspiration PrimeMUE · Traditional Siemens Medical Solutions USA, Inc.06/11/2013SE
K122510Icrco 3600MMUE · Traditional Icrco, Inc.04/26/2013SE
K122286Mammomal InspirationMUE · Traditional Siemens Medical Solutions USA, Inc.02/22/2013SE
K123995Microdose SiMUE · Special Philips Digital Mammography Sweden AB02/01/2013SE
K123873Contrast Enhanced Digital MammographyMUE · Traditional Hologic, Inc.01/29/2013SE
K121963Planmed Nuance DigiguideMUE · Traditional Planmed OY11/21/2012SE
K121674Aspire HD Plus, Aspire Hd-SMUE · Traditional Fujifilm Medical System U.S.A., Inc.09/21/2012SE
K120246Kodak Directview CR Mammography System (Using Carestream CR Mammography Cassette with…MUE · Traditional Carestream Health, Inc.08/28/2012SE
K120255Philips MicrodoseMUE · Traditional Philips Digital Mammography Sweden AB04/19/2012SE
K110717Konica Minolta Xpress Digital Mammography SystemMUE · Traditional Konica Minolta Medical & Graphic, Inc.12/23/2011SE
K111324CR Mammography System with Dx-M DigitizerMUE · Traditional Agfa Healthcare N.V.12/22/2011SE
K110798Senographe DS, Senographe EssentialMUE · Traditional General Electric Company12/22/2011SE
K111646Aspire CR for Mammography System (CRM); FCR Aspire CRN (Reader). Note for General…MUE · Traditional Fujifilm Medical System U.S.A., Inc.12/08/2011SE
K111434Giotto Image 3D, Giotto Image 3D-LMUE · Traditional Giotto USA, LLC10/27/2011SE
K103485Senograph DS, Senograph EssenstialMUE · Traditional General Electric Company10/06/2011SE
K111361Planmed Nuance ExcelMUE · Traditional Planmeca USA, Inc.09/23/2011SE
K110572Mammodiagnost DRMUE · Traditional Philips Medical Systems09/23/2011SE
K110729Aspire HD Full-Field Digital Mammography SystemMUE · Traditional Fujifilm Medical Systems U.S.A, Inc.09/01/2011SE
K110025Sectra Microdose MammographyMUE · Traditional Sectra Imtec AB04/28/2011SE

PMA approvals

Recalls

RecallProductFirmInitiated
Z-2167-2026 Class IISELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC…Complaints have been received of systems developing loose, missing, or broken internal bolts over time. Hologic, Inc04/08/2026Open, Classified
Z-0879-2026 Class IIMAMMOMAT Fusion;There were instances where the operator table was sold together with a bus-installation kit. The operator… Siemens Medical Solutions USA, Inc10/24/2025Open, Classified
Z-0878-2026 Class IIMAMMOMAT Inspiration;There were instances where the operator table was sold together with a bus-installation kit. The operator… Siemens Medical Solutions USA, Inc10/24/2025Open, Classified
Z-0877-2026 Class IIMAMMOMAT Revelation;There were instances where the operator table was sold together with a bus-installation kit. The operator… Siemens Medical Solutions USA, Inc10/24/2025Open, Classified
Z-0875-2026 Class IISenographe Pristina, Pristina Serena, and Pristina Serena 3D. for screening and…Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a… GE Medical Systems, LLC10/24/2025Open, Classified
Z-0862-2024 Class IISenographe PristinaX-ray exposure termination audible signal on Senographe Pristina with Efficiency Suite 2 intermittently not… GE Medical Systems, SCS12/26/2023Open, Classified
Z-2388-2020 Class IIMAMMOMAT Revelation, model no. 11343300 - Product Usage: is intended to be used for…If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy… Siemens Medical Solutions USA, Inc05/07/2020Terminated
Z-1343-2020 Class IISelenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions…The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing… Hologic, Inc.01/21/2020Terminated
Z-0160-2020 Class IIMammomat Revelation with software version VC10 and Biopsy Option Intended Use: For…An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may… Siemens Medical Solutions USA, Inc07/23/2019Terminated
Z-2132-2019 Class IIGE Pristina Serena - Product Usage: Pristina Serena is an optional accessory of…Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure. GE Healthcare, LLC05/16/2019Terminated
Z-2131-2019 Class IIGE Senographe Pristina - Product Usage: Senographe Pristina generates digital…Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure. GE Healthcare, LLC05/16/2019Terminated
Z-0685-2019 Class III-View Contrast Enhanced Digital Mammography, ASY-08109Calibration issue not possible to visualize contrast uptake in the subtracted mammography images Hologic, Inc.12/06/2018Terminated
Z-0662-2018 Class IIFuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed…FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The… Fujifilm Medical Systems U.S.A., Inc.08/04/2017Terminated
Z-0661-2018 Class IIASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE…FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The… Fujifilm Medical Systems U.S.A., Inc.08/04/2017Terminated
Z-0660-2018 Class IIAspire HD Plus, Aspire HD-s (for FDR-2000AWS) The Fujifilm Digital Mammography Systems…FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The… Fujifilm Medical Systems U.S.A., Inc.08/04/2017Terminated
Z-0659-2018 Class IIAspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR…FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The… Fujifilm Medical Systems U.S.A., Inc.08/04/2017Terminated
Z-0386-2018 Class IIASPIRE Cristalle (FDR MS-3500) Mammography machine The Fujifilm Digital Mammography…There was a case that the swivel arm went down unintentionally when it was moved up/down. Fujifilm Medical Systems U.S.A., Inc.07/14/2017Terminated
Z-0385-2018 Class IIAspire HD Plus (FDR MS-2500), also known as Aspire HD Plus F and Aspire HD-s (FDR…There was a case that the swivel arm went down unintentionally when it was moved up/down. Fujifilm Medical Systems U.S.A., Inc.07/14/2017Terminated
Z-1118-2017 Class IIMammomat Inspiration full, field digital,system,x-ray,mammographic Product Usage: The…Software error Siemens Medical Solutions USA, Inc01/11/2017Terminated
Z-2690-2016 Class IIFUJIFILM Digital Mammography System Aspire HD (FDR MS-1000) with Biopsy Positioner…If an exposure is aborted during a stereo biopsy workflow, an error may occur which would prevent the system… Fujifilm Medical Systems U.S.A., Inc.06/09/2016Terminated
Z-1585-2015 Class IIAspire HD, Full Field Digital, System, x-Ray, Mammographic; Model: FOR MS-1000. A…User Manual lacks a description for handling some errors even though the system can be recovered from such… Fujifilm Medical Systems U.S.A., Inc.02/03/2015Terminated
Z-0523-2015 Class IIMammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and…It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request)… Siemens Medical Solutions USA, Inc10/28/2014Terminated
Z-0086-2015 Class IISiemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is…The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch… Siemens Medical Solutions USA, Inc09/18/2014Terminated
Z-1700-2014 Class IIHologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030…Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the… Hologic, Inc.05/13/2014Terminated
Z-1638-2014 Class IIMAMMOMAT Inspiration. Intended for mammography exams, screening, diagnosis, and…There is a potential and possible hazard to the user when using the MAMMOMAT Inspiration PC monitor at the… Siemens Medical Solutions USA, Inc04/25/2014Terminated
Z-2301-2013 Class IISelenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The…Software: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most… Hologic, Inc.09/06/2013Terminated
Z-1787-2014 Class IIIBeekley Spots Light Image O-Spots for Mammography Catalog Number: 252 Product Usage: To…Light Image O-Spots Dispenser mislabeled as Soft n' Stretchy Light Image S-Spots Beekley Corporation06/19/2013Terminated
Z-1694-2013 Class IISiemens Mammomat Inspiration - full-field digital mammography Product Usage: MammographyFirm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that… Siemens Medical Solutions USA, Inc02/28/2013Terminated
Z-1519-2013 Class IIGE Healthcare Senographe Essential System. The Senographe 2000D system generates digital…GE Healthcare has recently become aware of a potential issue due to inaccessible Operator Instructions… GE Healthcare, LLC01/14/2013Terminated
Z-1642-2014 Class IIMXR-35 Diagnostic X-ray High Voltage Generator manufactured by DRGEM Corporation. x-ray…The MXR-35 X-ray Generator's labeling did not include the certification statement as required by 21 CFR Part… Fujifilm Medical System USA, Inc.06/29/2012Terminated
Z-1063-2012 GE SENOGRAPHE 2000D FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM ( SENO 2000D). The Senographe…GE Healthcare has become aware of inaccurate measurements on GE magnified images provided by Senographe… GE Healthcare, LLC10/17/2011Terminated
Z-0115-2012 GE SENOGRAPHE DS FULL FIELD DIGITAL ESSENTIAL MAMMOGRAPHY SYSTEM. The Senographe DS…GE Healthcare has become aware of inaccurate measurement on GE magnified images provided by Senographe DS… GE Healthcare, LLC09/22/2011Terminated
Z-0114-2012 GE SENOGRAPHE ESSENTIAL FULL FIELD MAMMOGRAPHY SYSTEM (SENO ESSENTIAL). The Senographe…GE Healthcare has become aware of inaccurate measurement on GE magnified images provided by Senographe DS… GE Healthcare, LLC09/22/2011Terminated
Z-0523-2012 MAMMOMAT Inspiration Full field digital mammographic x-ray systemLowering the swivel arm with "object table" while an operator's or patient's body parts are underneath the… Siemens Medical Solutions USA, Inc09/15/2011Terminated
Z-1041-2011 GE SENOGRAPHE ESSENTIAL FULL FIELD MAMMOGRAPHY SYSTEM (SENO ESSENTIAL) The Senographe…During installation it was discovered that the duplicate of the 21 CFR 1010 Certification label for the… GE Healthcare, LLC01/19/2011Terminated
Z-1044-2011 GE SENOGRAPHE DS FULL FIELD MAMMOGRAPHY SYSTEM MAMMOGRAPHIC FILM / SCREEN SYSTEMIt was noticed that 3 units received in Japan were missing the rating plate on the X-ray tube cover. There… GE Healthcare, LLC07/20/2010Terminated
Z-1043-2011 GE SENOGRAPHE ESSENTIAL FULL FIELD MAMMOGRAPHY SYSTEM ( SENO ESSENTIAL) The expected…It was noticed that 3 units received in Japan were missing the rating plate on the X-ray tube cover. There… GE Healthcare, LLC07/20/2010Terminated
Z-1233-2009 Siemens Mammomat Novation DR, full field digital mammography system. Model Number…Image may appear dark, requiring repeat image acquisition. Siemens Medical Solutions USA, Inc01/23/2009Terminated
Z-1546-2008 Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field…Marker Misalignment: The new software version contained a magnification factor that shows CAD markers… Lorad, A Hologic, Inc.02/28/2008Terminated