Amplitude-Integrated Electroencephalograph

FDA product code OMA · Class II (moderate risk) · 21 CFR 882.1400 · Neurology

Measure and record electrical activity of the brain by acquisition of amplitude-integrated electroencephalograph (electroencephalograph signals that have been filtered and displayed in a specific manner).

16510(k) clearances
0PMA approvals
3recalls

Manufacturers

510(k) clearances

510(k)DeviceApplicantDecision
K192889NeoOMA · Traditional Eemagine Medical Imaging Solutions GmbH01/30/2020SE
K163644Nihon Kohden QP-160AK EEG Trend ProgramOMA · Traditional Nihon Kohden Corporation05/19/2017SE
K161027Cadwell AmpliScanOMA · Traditional Cadwell Industries, Inc.11/08/2016SE
K152301Background Pattern Classification [BPc(TM)]OMA · Traditional Natus Medical Incorporated Dba Excel-Tech Ltd.…06/03/2016SE
K130238Nihon Kohden AE-918P Neuro UnitOMA · Traditional Nihon Kohden Corp.03/04/2015SE
K131789Cerebralogik- AeegOMA · Traditional Mennen Medical , Ltd.12/27/2013SE
K123079Olympic Brainz MonitorOMA · Traditional Excel-Tech Ltd. (Xltek)05/08/2013SE
K120485Nihon Kohden QP-160 Ak Trend Program with the Addition of Dsa Asymmetry Trend and FFT POMA · Special Nihon Kohden America, Inc.03/16/2012SE
K092573Nihon Kohden QP-160AK EEG Trend ProgramOMA · Traditional Nihon Kohden America, Inc.07/09/2010SE
K093949Olympic Brainz Monitor Model OBM70001OMA · Traditional Natus Medical, Inc.06/16/2010SE
K081551Twin Neurotrac-IiiOMA · Traditional Astro-Med, Inc.12/24/2008SE
K080217Component Neuromonitoring SystemOMA · Traditional Moberg Research, Inc.11/05/2008SE
K071449BRM3 Brain MonitorOMA · Traditional Brainz Instruments , Ltd.09/14/2007SE
K031149Olympic CFM 6000OMA · Special Olympic Medical Corp.05/14/2003SE
K983229Olympic Medical Lectromed Cerebral Function Monitor SystemOMA · Traditional Olympic Medical Corp.03/23/1999SE
K791580Cerebral Function Monitor 870OMA · Traditional Applied Medical Research09/04/1979SE

Recalls

RecallProductFirmInitiated
Z-1925-2025 Class IIBrand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB…Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately… Natus Neurology DBA Excel Tech., Ltd. (XLTEK)05/07/2025Open, Classified
Z-1338-2023 Class IIMoberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term…Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact… Moberg Research, Inc.02/20/2023Open, Classified
Z-1630-2020 Class IInatus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalographThe OBM system functional test (impedance and noise) was not carried out on the affected items prior to… Natus Neurology DBA Excel Tech., Ltd. (XLTEK)03/05/2020Terminated