Recall Z-0004-2011

Verathon, Inc. · initiated August 23, 2010

Terminated

Product

AortaScan AMI 9700 The AortaScan AMI 9700 is an ultrasound device that projects ultrasound energy into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurements

Codes / lots: Serial numbers: A7001001, A7001005, A7001040, A7001041, A7001042, A7001043, A7001044, A7001045, A7001046, A7001047, A7001048, A7001049, A7001050, A7001051, A7001052, A7001053, A7001054, A7001055, A7001056, A7001057, A7001062, A7001065, A7001066, A7001067, A7001069, A7001071, A7001089, A7001090, A7001091, A7001092, A7001093, A7001094, A7001095, A7001096, A7001097, A7001098, A7001099, A7001100, A7001101, A7001102, A7001103, A7001106, A7001107, A7001108, A7001109, A7001110, A7001111, A7001112, A7001113, A7001114, A7001115, A7001116, A7001117, A7001118, A7001119, A7001120, A7001121, A7001122, A7001123, A7001124, A7001125, A7001127, A7001128, A7001129, A7001130, A7001131, A7001132, A7001133, A7001134, A7001135, A7001136, A7001138, and A7001139.

Reason for recall

The AortaScan AMI 9700 devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.

FDA-determined cause: Software design

Action taken

Verathon sent an URGENT Medical Device Recall Notification letter dated August 12, 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Verathon provided an addendum to the User Manual with instructions on how to reset the instrument. Customers were advised to: " Add the enclosed addendum to their current Users Manual. " Update the application software of the device to v2.7.0 or higher. If the customers had questions or needed support for the software update process, they were to contact Verathon Medical Customer Care at (425) 867-1348 or (800) 331-2313 or cservice@verathon.com.

Details

Recall number
Z-0004-2011
Initiated
August 23, 2010
Posted by FDA
October 4, 2010
Terminated
September 8, 2011
Product code
IYO – System, Imaging, Pulsed Echo, Ultrasonic
Quantity
73 units
Distribution
Worldwide Distribution - USA and the countries of Australia, Canada, China, Colombia, European Union, United Kingdom, Saudi Arabia, and Singapore.
Recalling firm
Verathon, Inc., Bothell, WA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0003-2011 BVI 9600 Bladder Volume Instrument, Aorta Scan Mode The BladderScan BVI 9600 with…The BladderScan BVI 9600 with AortaScan devices may experience the loss of a factory installed software… 08/23/2010Terminated
Z-0005-2011 BladderScan BVM 9500 The BladderScan BVM 9500 is intended to project ultrasound energy…The BladderScan BVM 9500 devices may experience the loss of a factory installed software configuration file… 08/23/2010Terminated