Recall Z-0011-2009

Roche Diagnostics Corp. · initiated June 11, 2008

Terminated

Product

Roche Diagnostics ALBT2, Tina-quant Albumin Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04469658190. An in-vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of albumin concentration in serum and plasma.

Codes / lots: Lot 60087801.

Reason for recall

Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.

FDA-determined cause: Labeling Change Control

Action taken

Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.

Details

Recall number
Z-0011-2009
Initiated
June 11, 2008
Posted by FDA
October 9, 2008
Terminated
September 22, 2009
Product code
CIX – Bromcresol Green Dye-Binding, Albumin
Quantity
124
Distribution
Nationwide Distribution
Recalling firm
Roche Diagnostics Corp., Indianapolis, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0006-2009 Roche Diagnostics LDL_C, LDL- Cholesterol Plus 2nd generation, COBAS INTEGRA, cobas c…Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. 06/11/2008Terminated
Z-0007-2009 Roche Diagnostics IRON2, Iron Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics…Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. 06/11/2008Terminated
Z-0008-2009 Roche Diagnostics CREP2, Creatinine Plus Ver. 2, COBAS INTEGRA, cobas c systems, Roche…Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. 06/11/2008Terminated
Z-0009-2009 Roche Diagnostics STFR Tina-quant Soluble Transferrin Receptor, COBAS INTEGRA, cobas c…Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. 06/11/2008Terminated
Z-0010-2009 Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c…Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. 06/11/2008Terminated
Z-0012-2009 Roche Diagnostics NAPA2, N-Acetyl-Procainamide, COBAS INTEGRA, Roche Diagnostics…Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. 06/11/2008Terminated