Recall Z-0012-2020

Inter-Med Llc · initiated June 27, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Vista Dental Products Sodium Hypochlorite - 6%, 10 - 12cc Pre-filled Syringes, REF 502300

Codes / lots: Lot/Work Order Numbers: 2018-1348 2018-1585 2018-1586 2018-1852 20182106, UDI (01)10818207020465 20182302, UDI (01)10818207020465 20182458, UDI (01)10818207020465 20190725, UDI (01)10818207020465 20190835, UDI (01)10818207020465

Reason for recall

The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

FDA-determined cause: Process control

Action taken

The firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.

Details

Recall number
Z-0012-2020
Initiated
June 27, 2019
Posted by FDA
October 1, 2019
Terminated
August 10, 2021
Product code
KJJ – Cleanser, Root Canal
Quantity
1992 units
Distribution
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
Recalling firm
Inter-Med Llc, Racine, WI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0011-2020 Class IIVista Dental Products CanalClean Endodontic Irrigation Kit, (1) 12mL - Chlor-XTRA, 6mL -…The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of… 06/27/2019Terminated
Z-0013-2020 Class IIVista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of… 06/27/2019Terminated
Z-0014-2020 Class IIVista Dental Products Sodium Hypochlorite - 6%, QTY: 12, 3mL Pre-filled Syringes, REF…The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of… 06/27/2019Terminated
Z-0015-2020 Class IIVista Dental Products 3% Sodium Hypochlorite, 10- 3cc Pre-filled Syringes, REF 502355The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of… 06/27/2019Terminated
Z-0016-2020 Class IIVista Dental Products Chlor-XTRA - 3mL Prefilled Syringes, QTY: 12, REF 503825The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of… 06/27/2019Terminated
Z-0017-2020 Class IIVista Dental Products Chlor-XTRA - 12mL Prefilled Syringes, QTY: 8, REF 503850The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of… 06/27/2019Terminated