Recall Z-0028-2012

Beckman Coulter Inc. · initiated August 12, 2010

Terminated

Product

Coulter PrepPlus 2, Part Number: 378600, Serial Numbers: 1001008698, 1001008699, and 10018700. Coulter PrePlus 2 is a microprocessor-controlled pipetting and diluting system, designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.

Codes / lots: Serial Numbers: 1001008698, 1001008699, and 10018700.

Reason for recall

The recall was initiated because Beckman Coulter has confirmed an issue with the Coulter PrePlus 2. The reagent level sensing is impaired and will give no warning when a reagent vial is low or out of reagent. Due to the size difference in the vertical mounts, the reagent rack is not aligned optimally for the probe to aspirate or level sense the antibody. This misalignment yields impaired reagen

FDA-determined cause: Mixed-up of materials/components

Action taken

Beckman Coulter initiated an URGENT: PRODUCT CORRECTIVE ACTION (PCA) letter dated August 24, 2010, with attached PCA Response Form to all affected customers who purchased the Coulter PrepPlus 2. The letter provided the customers with an explanation of the problem identified and an action to be taken. The customers were instructed to not use the PrepPlus2 until their Beckman Coulter representative corrects the problem. A review of all results generated from samples prepared from the system should be conducted. Any unexpected dim fluorescence and/or decreased number stained cells should be evaluated for re-analysis, including preparing a new sample. Customers were also instructed to complete and return the enclosed response form within 10 days. Customers with any questions regarding this PCA, were instructed to call Beckman Coulter Customer service at (800) 526-7694. Outside of USA, customers were instructed to contact their local Beckman Coulter Representative.

Details

Recall number
Z-0028-2012
Initiated
August 12, 2010
Posted by FDA
October 7, 2011
Terminated
April 4, 2012
Product code
JQW – Station, Pipetting And Diluting, For Clinical Use
Quantity
391 units total (193 in US)
Distribution
Worldwide distribution.
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record