Recall Z-0040-2024

AB Sciex, LLC · initiated August 30, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Codes / lots: UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203

Reason for recall

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

FDA-determined cause: Component design/selection

Action taken

An URGENT: RUO and MEDICAL DEVICE FIELD ACTION notification letter dated 7/20/23 was sent to customers. Actions to be taken by the Customer/User: SCIEX will provide a replacement nut for eligible sources and systems. Instructions will be provided to guide you through the replacement of the nut. The replacement parts with installation instructions are estimated to begin shipping in October 2023. SCIEX is currently notifying affected customers of the upcoming countermeasure. Upon availability, SCIEX will send you the replacement nut with instructions on how to install it on the ion source probe. We sincerely apologize for any inconvenience that this may cause you. SCIEX aims to provide you with products of the highest quality. If you have any questions regarding this notice, please contact https://sciex.com/aboutus/ contact-us or https://sciex.com/support/request-support.

Details

Recall number
Z-0040-2024
Initiated
August 30, 2023
Posted by FDA
October 5, 2023
Product code
DOP – Mass Spectrometer, Clinical Use
Quantity
14
Distribution
US Nationwide.
Recalling firm
AB Sciex, LLC, Framingham, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0037-2024 Class IIAPI3200MD Mass Spectrometer, Part Number 4466230In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an… 08/30/2023Open, Classified
Z-0038-2024 Class II4500MD Triple Quad Mass Spectrometer, Part Number 5032522In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an… 08/30/2023Open, Classified
Z-0039-2024 Class IICitrine QTRAP Mass Spectrometer, Part Number 5063685In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an… 08/30/2023Open, Classified