Recall Z-0048-2022
ZOLL Circulation, Inc. · initiated August 6, 2021
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Zoll REF: 8700-0783-01 (IC-4593), QUATTRO Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin, Sterile EO, Rx Only, UDI: (01)00849111075282
Codes / lots: All Lots
Reason for recall
Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing
FDA-determined cause: Labeling design
Action taken
Between the dates of August 27, 2021 and September 3, 2021, ZOLL, issued an "Urgent Medical Device Recall (Correction)" Notification to all affected US consignees via UPS. In addition to informing consignees about the labeling correction, Zoll asked consignees to take the following actions: 1. If you have any questions, please contact Doug Lam (Director of Quality and Compliance) or Sam Nanavati (VP of Quality and Regulatory Affairs) at IVTM_Recall@zoll.com. 2 Provide this notice (letter) to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. 3 Maintain a copy of the recall notification and attachment with the IFU for reference. ZOLL will provide a method to obtain the updated IFU once it is released. 4 If you experience any adverse reactions or quality problems with the use of our products (catheters), you may report adverse reactions or quality problems to the FDA s MedWatch Adverse Event Reporting program either online (https://www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=consumer.reporting1), by regular mail or by fax (1-800-332-0178). 5 Sign and return the pre-stamped post card (enclosed) acknowledging that you have read and understood the content and requirements on this recall notice (letter) within 5 working days from receipt of this notice. 6. If you do not believe that you are the right person to implement the above-mentioned actions, please forward this letter to the right person in your organization. Actions Underway by ZOLL: 1. ZOLL will submit a revised IFU to FDA with heavy focus on providing clear instructions to the users on how to investigate/mitigate a potential catheter leak whenever a depleted saline bag is noticed by the users. 2. ZOLL will provide all affected users a method to obtain the revised IFU once it is reviewed and cleared by the FDA. 3. ZOLL is retraining all Sales staff to the revised IFU. 4. If we could be of any further assi
Details
- Recall number
- Z-0048-2022
- Initiated
- August 6, 2021
- Terminated
- October 18, 2024
- Product code
- NCX – System, Hypothermia, Intravenous, Cooling
- Quantity
- 20,475 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries Australia, CANADA, CHINA, GERMANY, HUNGARY, PORTUGAL, SINGAPORE, TAIWAN, THAILAND.
- Recalling firm
- ZOLL Circulation, Inc., San Jose, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0039-2022 Class II | Zoll REF: 8700-0782-01 (IC-3893), ICY Intravascular Heat Exchange Catheter Kit, with…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0040-2022 Class II | Zoll, REF: 8700-0791-01 (IC-4593), Premium Access Kit QUATTRO, Intravascular Heat…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0041-2022 Class II | ZOLL, REF: 8700-0781-01 (CL-2295), COOL LINE, Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0042-2022 Class II | ZOLL, REF 8700-0793-01 (SL-2593) Premium Access Kit SOLEX 7, Intravascular Heat Exchange…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0043-2022 Class II | ZOLL, REF 8700-0793-40 (SL-2593AE) SOLEX 7, Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0044-2022 Class II | ZOLL, REF 8700-0781-14 (CL-2295CO), COOL LINE Intravascular Heat Exchange Catheter kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0045-2022 Class II | ZOLL, REF 8700-0781-40 (CL-2295AE), COOL LINE Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0046-2022 Class II | ZOLL, REF 8700-0787-40 (IC-3893AE), ICY Intravascular Heat Exchange Catheter Kit with…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0047-2022 Class II | Zoll REF: 8700-0782-40 (IC-3893AE), ICY Intravascular Heat Exchange Catheter Kit with…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0049-2022 Class II | Zoll REF: 8700-0783-40 (IC-4593AE), QUATTRO Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0050-2022 Class II | Zoll REF: 8700-0782-14 (IC-3893CO), ICY Intravascular Heat Exchange Catheter Kit with…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0051-2022 Class II | Zoll REF: 8700-0783-14 (IC-4593CO), QUATTRO Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0052-2022 Class II | Zoll REF: 8700-0793-14 (SL-2593CO), SOLEX 7 Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0053-2022 Class II | Zoll, REF: 8700-0783-03 (IC-4593AE) QUATTRO Intravascular Heat Exchange Catheter Kit…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0054-2022 Class II | Zoll, REF 8700-0781-03 (CL-2295AE), COOLINE Intravascular Heat Intravascular Heat…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0055-2022 Class II | Zoll, REF 8700-0782-03 (IC-3893AE), ICY Intravascular Heat Intravascular Heat Exchange…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0056-2022 Class II | Zoll, REF 8700-0695-40 (IC-3893AE), ICY Intravascular Heat Intravascular Heat Exchange…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0057-2022 Class II | Zoll, REF 8700-0657-40 (IC-3893AE), ICY Intravascular Heat Intravascular Heat Exchange…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0058-2022 Class II | Zoll, REF 8700-000766-01, Quattro IVTM Premium Disposable Pack and Start UP Kit (6 ft.)Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0059-2022 Class II | Zoll, REF 8700-000865-01, Cool Line IVTM Disposable Pack and Start UP Kit (6 ft.)Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0060-2022 Class II | Zoll, REF 8700-000866-01, ICY IVTM Disposable Pack and Start UP Kit (6 ft.)Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0061-2022 Class II | Zoll, REF 8700-000867-01, Quattro IVTM Disposable Pack and Start UP Kit (6 ft.)Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0062-2022 Class II | Zoll, REF 8700-000868-01, Solex 7 IVTM Disposable Pack and Start UP Kit (6 ft.)Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0063-2022 Class II | Zoll, REF 8700-000869-40, Cool Line (Heparin) IVTM Disposable Pack and Start UP Kit (6…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0064-2022 Class II | Zoll, REF 8700-000870-40, ICY (Heparin) IVTM Disposable Pack and Start UP Kit (6 ft.)…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0065-2022 Class II | Zoll, REF 8700-000871-40, Quattro (Heparin) IVTM Disposable Pack and Start UP Kit (6…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |
| Z-0066-2022 Class II | Zoll, REF 8700-000872-40, SOLEX 7(Heparin) IVTM Disposable Pack and Start UP Kit (6…Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing | 08/06/2021Terminated |