Recall Z-0050-2021

Ra Medical Systems, Inc. · initiated July 24, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in ablating a channel in occlusive peripheral vascular disease.

Codes / lots: Model Number DABRA Excimer Laser RA-308; Serial Numbers: RA00027, RA00039, RA00046, RA00049, RA00050, RA00054, RA00055, RA00056, RA00057, RA00058, RA00063, RA00064, RA00065, RA00067, RA00069, RA00070, RA00073, RA00074, RA00081, RA00082, RA00083, RA00084, RA00085, RA00086, RA00088, RA00089, RA00097, RA00098, RA00103, RA00114, RA00115, RA00117, RA00118, RA00120, RA00121, RA00124, RA00125, RA00127, RA00135, RA00136, RA00137, RA00138, RA00143, RA00147, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.

Reason for recall

Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.

FDA-determined cause: Device Design

Action taken

On August 10, 2020, the firm sent an "Urgent Medical Correction" letter to customers, informing them of the incorrect laser wheel size, which upon moving or adjusting the laser with the wheels in locked position may fall over. If this were to happen during a medical procedure, user or patient injury may occur.

Details

Recall number
Z-0050-2021
Initiated
July 24, 2020
Terminated
August 9, 2023
Product code
PDU – Catheter For Crossing Total Occlusions
Quantity
54 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, FL, ID, IL, IN, KS, KY, LA, MA, MD, MI, NC, NV, NY, OR, OK, PA, TX, WA and the country of Japan.
Recalling firm
Ra Medical Systems, Inc., Carlsbad, CA

Official FDA recall record