Recall Z-0051-2014

Synergetics Inc · initiated July 30, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Synergetics 25 gauge Soft Tip Cannula, packaged in 6 units per box. Catalog number: 30.08.25. Single-use 25 gauge cannula used in vitreoretinal surgery.

Codes / lots: Lot number: M222240

Reason for recall

One lot of 25 gauge Soft-Tip Disposable Cannulas are co-mingled with 23 gauge cannulas; the product is labeled as 25 gauge cannulas.

FDA-determined cause: Labeling Change Control

Action taken

The firm sent recall notification letters to their consignees on 7/30/13.

Details

Recall number
Z-0051-2014
Initiated
July 30, 2013
Posted by FDA
October 22, 2013
Terminated
April 7, 2014
Product code
HMX – Cannula, Ophthalmic
Quantity
48/6-cannula boxes
Distribution
Nationwide Distribution-USA including the states of MO, CO, KY, TN, NC, PA, IL CA, NY, IN, MI, LA, NJ and OH.
Recalling firm
Synergetics Inc, O Fallon, MO