Recall Z-0054-2012

Terumo Cardiovascular Systems Corporation · initiated September 13, 2011

Terminated

Product

SARNS INTRACARDIAC SUCKER, BULK, NON-STERILE

Codes / lots: Catalog number 11009010

Reason for recall

Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker.

FDA-determined cause: Nonconforming Material/Component

Action taken

On 9/13/2011 Terumo sent a letter entitled "URGENT MEDICAL DEVICE REMOVAL" to its customers that included the name and description of the device, lot numbers, the potential hazard, correction, the reason for the recall and instructions. Customers are to review the notice, ensure that all users are aware of the issue, and a recall response form to be completed and returned. Customers are to call Terumo CVS Customer Service to obtain a Return Goods Authorization number. The letter also included MedWatch reporting information and stated that Customers with questions should call 1-800-521-2818.

Details

Recall number
Z-0054-2012
Initiated
September 13, 2011
Posted by FDA
October 17, 2011
Terminated
December 10, 2012
Product code
DTS – Sucker, Cardiotomy Return, Cardiopulmonary Bypass
Quantity
5700
Distribution
Worldwide Distribution -- USA, Belgium, Australia, Mexico, and Japan.
Recalling firm
Terumo Cardiovascular Systems Corporation, Ann Arbor, MI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0053-2012 Sarns Rigid Intracardiac Sucker. The subject device is a single use, sterile device with…Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the… 09/13/2011Terminated