Recall Z-0057-2008

I-Flow Corporation · initiated March 27, 2007

Terminated

Product

ON-Q PainBuster with OnDemand (270 ml, 2 ml/hr + 5 ml bolus /60 min refill), Model PMB01, I-Flow Corporation, Lake Forest, CA 92630 U.S.A.

Codes / lots: 6A4171 (exp. MAR 2009), 694133 (exp. FEB 2009), 654920 (exp. OCT 2008), 5A4732 (exp. MAR 2008), 584699 (exp. JAN 2008)

Reason for recall

Misassembly: The ON-Q PainBuster PMB01 package may contain a Soaker catheter instead of a Non-Soaker catheter. As a result, the bolus button may remain latched in the down position which may cause the the device to deliver continuously at 7 ml/hr instead of 2 ml/hr.

FDA-determined cause: Process control

Action taken

An Important Recall Notification Letter was faxed to affected customers on March 27, 2007. Affected customers were notified of the recall, the reason for recall, and instructed to immediately quarantine any remaining inventory. A Customer Response (and Fax back) form was included with the Recall Notification. Affected customers were requested to return the FAX back form to I-Flow within 48 hours of receipt. Replacement product will be sent out as applicable.

Details

Recall number
Z-0057-2008
Initiated
March 27, 2007
Posted by FDA
October 6, 2007
Terminated
October 7, 2009
Product code
MED – Sterilant, Medical Devices
Quantity
2,587 units
Distribution
Worldwide, USA and Australia
Recalling firm
I-Flow Corporation, Lake Forest, CA

Official FDA recall record