Recall Z-0068-2009
Bard Peripheral Vascular Inc · initiated August 29, 2008
Product
Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards Lifesciences LLC Irvine, CA 92614, Model EX090401CD. The LifeStent FlexStar self-expanding biliary stent system is intended for use in the palliation of malignant strictures (neoplasms) in the biliary tree.
Codes / lots: Lot Numbers: FR6H0142, FR6J0246, FR6K0321, and R58301781
Reason for recall
Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible.
FDA-determined cause: Process control
Action taken
The customer letters for the tip to sheath gap were sent on 8/29/2008 with return receipt. The letter advises that C.R. Bard has determined that some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible. Per the letter, the firm's analysis has determined that a gap between the tip of the delivery system and the delivery system sheath may be associated with difficult advancement to the treatment site. Users are instructed to inspect all devices to verify that there is not a gap between the tip and sheath (orange colored guidewire lumen is visible), prior to use and to not use any devices that exhibit a tip to sheath gap condition. If devices that exhibit a tip to sheath gap are identified, customers are advised to call Customer Service at 1-800-321-4254 to return the device and a replacement will be provided at no cost. The letter also advises that the Instructions for Use for the LifeStent FlexStar System will be updated to include the inspection step which is provided in the letter.
Details
- Recall number
- Z-0068-2009
- Initiated
- August 29, 2008
- Posted by FDA
- October 10, 2008
- Terminated
- April 6, 2012
- Product code
- FGE – Stents, Drains And Dilators For The Biliary Ducts
- Quantity
- 209 units
- Distribution
- Nationwide
- Recalling firm
- Bard Peripheral Vascular Inc, Tempe, AZ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0035-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0036-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0037-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0038-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0039-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0040-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0041-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0042-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0043-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0044-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0045-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0046-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0047-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0048-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0049-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0050-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0051-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0052-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0053-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0054-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0055-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0056-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0057-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0058-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0059-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0060-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0061-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0062-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0063-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |
| Z-0064-2009 | Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards…Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary… | 08/29/2008Terminated |