Recall Z-0073-2025

Merit Medical Systems, Inc. · initiated August 19, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

StabiliT TOUCH Syringe, REF: IN8VCF/B

Codes / lots: REF/UDI-DI/Lot(Expiration): IN8VCF/B/00884450662144/H2918440(5/31/2027), H2938015(7/26/2027).

Reason for recall

Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.

FDA-determined cause: Packaging process control

Action taken

On 8/19/24, recall notices were emailed to customers asking them to do the following: 1) If affected devices are discovered, quarantine them, discontinue use and distribution, and return to the recalling firm. 2) Ensure that applicable personnel within your organization are made aware of this field action. 3) If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4) Complete and return the customer response form via email to response@merit.com If you have any questions concerning this communication, please contact the firm's Sales Representatives or Customer Service via email at RESPONSE@merit.com or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri.

Details

Recall number
Z-0073-2025
Initiated
August 19, 2024
Posted by FDA
October 11, 2024
Product code
OAR – Injector, Vertebroplasty (Does Not Contain Cement)
Quantity
202
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of France, Italy, Jamaica, Ireland, Paraguay, Australia, India, Nepal, Japan.
Recalling firm
Merit Medical Systems, Inc., South Jordan, UT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0064-2025 Class IIBlue Diamond Digital Inflation Device, REF: IN7152/C, IN7352/C, IN7403/C, IN7802/C…Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0065-2025 Class IIbasixTOUCH Inflation Device, REF: IN8140/AInflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0066-2025 Class IIDiamondTOUCH Digital Inflation Device, REF: IN9135/A, IN9152/AInflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0067-2025 Class IIIntelliSystem Inflation Device, REF: IN1525/DInflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0068-2025 Class IIBlueFIRE Inflation Device, REF: IN10140Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0069-2025 Class IIMONARCH Inflation Device, REF: IN2430/B, IN2530/B, IN2130/BInflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0070-2025 Class IICustom Inflation Kit, REF: K05-00052M, K05-00053H, K05-00109G, K05-00137K, K05-00296L…Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0071-2025 Class IIbasixSKY Inflation Device, REF: SKY1802Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified
Z-0072-2025 Class IIbasixTOUCH Inflation Device, REF: IN8130/JPC, IN8152/C, IN8302/C, IN8352/C, IN8802/C…Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may… 08/19/2024Open, Classified