Recall Z-0091-2015

Codman & Shurtleff, Inc. · initiated September 15, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated

Codes / lots: All lots with expiration on or before August 2017 (2017-08)

Reason for recall

tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)

FDA-determined cause: Device Design

Action taken

Codman Neuro sent an Urgent Notice dated September 12, 2014, to all affected customers via overnight mailing and OUS customers notified via email the week of September 15, 2014. Accounts were requested to isolate all inventory of the affected product and return to Stericycle at the following address: Stericycle 2670 Executive Drive, Suite A Attn: Event #8894 Indianapolis, IN 46241 For patients currently being managed with the EDS 3 System, the system should be replaced immediately. In the case where no substitute drainage system is immediately available, the EDS 3 System may continue to be used until an alternative product can be obtained. If you have any questions or concerns regarding this notification, please contact your local Codman Neuro Representative or Scientific and Medical Affairs at SciMedAffairs@its.jnj.com or (866) 685-7325. For questions regarding this recall call 508-880-8000.

Details

Recall number
Z-0091-2015
Initiated
September 15, 2014
Posted by FDA
October 20, 2014
Terminated
August 11, 2017
Product code
JXG – Shunt, Central Nervous System And Components
Quantity
1145 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to: United Arab Emirates Argentina Austria Australia Aruba Belgium Bahrain Bolivia Brazil Bahamas Canada Switzerland Chile People's Republic of China Columbia Costa Rica Cyprus Czech Republic Germany Algeria Estonia Egypt Spain Finland France Gabon Great Britain French Guiana Greece Guatemala Hong Kong Croatia Hungary Ireland Israel…
Recalling firm
Codman & Shurtleff, Inc., Raynham, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0089-2015 Class IICodman EDS 3 Drainage System with Ventricular Catheter (External Drainage System)…tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or… 09/15/2014Terminated
Z-0090-2015 Class IICodman EDS 3 Drainage System (no Ventricular Catheter) (External Drainage System)…tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or… 09/15/2014Terminated