Recall Z-0097-2007

Conmed Corporation · initiated October 12, 2006

Terminated

Product

ConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-002. --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.

Codes / lots: REF/Product Code 60-6045-002, all lot codes manufactured between 08/25/2003 and 03/24/2005. [Lot codes on unopened boxes and packaging contain a lot code in the following form: for example 0409271 - 04 represents the year 2004, 09 represents the month of the year, 27 represents the day of the month, and the last digit 1 represents the manufacturing shift code. Lot codes are located on the device in this form: for example 40927 - 4 represents the year 2004, 09 represents the month of the year, and 27 represents the day of the month.]

Reason for recall

The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.

FDA-determined cause: Nonconforming Material/Component

Action taken

On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.

Details

Recall number
Z-0097-2007
Initiated
October 12, 2006
Posted by FDA
October 27, 2006
Terminated
August 30, 2012
Product code
GCJ – Laparoscope, General & Plastic Surgery
Quantity
83 units
Distribution
Worldwide
Recalling firm
Conmed Corporation, Utica, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0083-2007 ConMed DetachaTip Laparoscopic Instrument: Allis Multiple Use Grasper, 5mm x 33cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0084-2007 ConMed DetachaTip Laparoscopic Instrument: Allis Multiple Use Grasper, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0085-2007 ConMed DetachaTip Laparoscopic Instrument: Babcock Multiple Use Grasper, 5mm x 33cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0086-2007 ConMed DetachaTip Laparoscopic Instrument: Babcock Multiple Use Grasper, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0087-2007 ConMed DetachaTip Laparoscopic Instrument: Curved ''Maryland'' Multiple Use Grasper, 5mm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0088-2007 ConMed DetachaTip Laparoscopic Instrument: Curved ''Maryland'' Multiple Use Grasper, 5mm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0089-2007 ConMed DetachaTip Laparoscopic Instrument: Endoweave Multiple Use Grasper, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0090-2007 ConMed DetachaTip Laparoscopic Instrument: Fenestrated (Duckbill) Multiple Use Grasper…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0091-2007 ConMed DetachaTip Laparoscopic Instrument: Meeker (Right Angle) Multiple Use Grasper…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0092-2007 ConMed DetachaTip Laparoscopic Instrument: Meeker (Right Angle) Multiple Use Grasper…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0093-2007 ConMed Laparoscopic Instrument: Allis Disposable Grasper, 5mm x 33cm length, REF/Product…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0094-2007 ConMed Laparoscopic Instrument: Allis Disposable Grasper, 5mm x 43cm length, REF/Product…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0095-2007 ConMed Laparoscopic Instrument: Babcock Disposable Grasper, 5mm x 33cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0096-2007 ConMed Laparoscopic Instrument: Babcock Disposable Grasper, 5mm x 43cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0098-2007 ConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Grasper, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0099-2007 ConMed Laparoscopic Instrument: Endoweave Disposable Grasper, 5mm x 33cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0100-2007 ConMed Laparoscopic Instrument: Endoweave Disposable Grasper, 5mm x 43cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0101-2007 ConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Grasper, 5mm x 33cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0102-2007 ConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Grasper, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0103-2007 ConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Grasper, 5mm x 33cm…The company was made aware of instances where the grasper jaws broke during laparoscopic procedures. When… 10/12/2006Terminated
Z-0104-2007 ConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Grasper, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0105-2007 ConMed Laparoscopic Instrument: Allis Disposable Dissector, 5mm x 33cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0106-2007 ConMed Laparoscopic Instrument: Allis Disposable Dissector, 5mm x 43cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0107-2007 ConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Dissector, 5mm x 33cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0108-2007 ConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Dissector, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0109-2007 ConMed Laparoscopic Instrument: Endoweave Disposable Dissector, 5mm x 33cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0110-2007 ConMed Laparoscopic Instrument: Endoweave Disposable Dissector, 5mm x 43cm length…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0111-2007 ConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Dissector, 5mm x 33cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0112-2007 ConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Dissector, 5mm x 43cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated
Z-0113-2007 ConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Dissector, 5mm x 33cm…The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube. 10/12/2006Terminated