Recall Z-0100-2017

Roche Diagnostics Operations, Inc. · initiated July 15, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Roche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612 is a computer-controlled stand-alone system for aliquoting and sorting of barcoded sample tubes. The system operates with an online connection to the host.

Codes / lots: Not applicable

Reason for recall

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

FDA-determined cause: Device Design

Action taken

On 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum.

Details

Recall number
Z-0100-2017
Initiated
July 15, 2016
Terminated
June 5, 2017
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
9014 in total
Distribution
US Nationwide Distribution
Recalling firm
Roche Diagnostics Operations, Inc., Indianapolis, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0094-2017 Class IIRoche/Hitachi MODULAR Analytics Combination Systems Modular D/P: Clinical Chemistry…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0095-2017 Class IIcobas e411 Immunoassay Analyzer e411: Immunoassay Analyzer Product Usage: The cobas…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0096-2017 Class IIcobas 6000 Series system c6000, Chemistry (Photometric, Discrete), for clinical Use.…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0097-2017 Class IIcobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0098-2017 Class IIModular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0099-2017 Class IIURISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0101-2017 Class IIcobas p 701 Post-Analytical Unit calculator/data processing module, for clinical use…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0102-2017 Class IIcobas p 501 Post-Analytical Unit calculator/data processing module, for clinical use…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0103-2017 Class IIVS II Aliquoter System calculator/data processing module, for clinical use Product…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0104-2017 Class IIRSD 800A Aliquoting System calculator/data processing module, for clinical use Product…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0105-2017 Class IIRSD 800 Sorting System calculator/data processing module, for clinical use Product…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0106-2017 Class IIcobas p 612 Pre-Analytical Instrument calculator/data processing module, for clinical…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0107-2017 Class IIcobas p 512 Pre-Analytical Instrument calculator/data processing module, for clinical…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0108-2017 Class IIcobas p 471 centrifuge unit calculator/data processing module, for clinical use Product…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0109-2017 Class IIcobas p 312 Pre-Analytical System calculator/data processing module, for clinical use…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated
Z-0110-2017 Class IIcobas 6800/8800 system Hepatitis Viral B DNA Detection Product Usage: The cobas p512…Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers… 07/15/2016Terminated