Recall Z-0101-2015
Estill Medical Technologies, Inc · initiated July 25, 2014
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.
Codes / lots: 06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014
Reason for recall
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
FDA-determined cause: Device Design
Action taken
Thermal Angel sent a Notice of Corrections letter dated July 25, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to communicate the contents of the notice to all employees, consignees, consultants, or other parties affected by this correction. The firm also updated the labeling and instructions for use. For additional information cusotmers were instructed to contact Medical Technologies at (877)354-0286.The recalling firm issued a Notice of Corrections to its distributors and customers. For questions regarding this recall call 214-561-6001.
Details
- Recall number
- Z-0101-2015
- Initiated
- July 25, 2014
- Posted by FDA
- October 21, 2014
- Terminated
- August 21, 2015
- Product code
- BSB – Warmer, Blood, Non-Electromagnetic Radiation
- Quantity
- 6,886
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Serbia.
- Recalling firm
- Estill Medical Technologies, Inc, Dallas, TX