Recall Z-0105-2021
Cook Inc. · initiated August 31, 2020
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.
Codes / lots: All lots
Reason for recall
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
FDA-determined cause: No Marketing Application
Action taken
URGENT: MEDICAL DEVICE CORRECTION notification letters dated 8/31/20 were sent to customers. Actions to be Taken by the Customer 1. If affected products have already been placed within a patient, the indwell time should not exceed 4 weeks. If the current indwell time has exceeded 4 weeks, consider removing and/or replacing the product. 2. Understand that if an affected product is placed within a patient longer than 4 weeks, there is an increased potential for breakage, kinking, encrustation, migration, and/or infection to occur. 3. Please maintain a copy of this notice with the current IFU or product(s) in your inventory. 4. If you have affected product in inventory, you may continue to use these products. Cook Medical is not requesting product be returned. 5. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 6. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 7. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is initiating a voluntary correction to inform customers that the indicated indwell time is changing for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set. The current IFU is in process of being updated and will be provided with orders placed following implementation. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your
Details
- Recall number
- Z-0105-2021
- Initiated
- August 31, 2020
- Terminated
- July 1, 2021
- Product code
- LJE – Catheter, Nephrostomy
- Quantity
- 368
- Distribution
- Worldwide distribution - US Nationwide.
- Recalling firm
- Cook Inc., Bloomington, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0106-2021 Class II | Pediatric Nephrostomy Stent Set, Reference Part Numbers 080208-S6 and 080208-S7, Order…The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set… | 08/31/2020Terminated |