Recall Z-0117-2023

Abbott Vascular · initiated September 8, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT, TriClip G4 NT Delivery System CLIN/TCDS0303-NT, TriClip G4 NTW Delivery System CLIN/TCDS0303-NTW, TriClip G4 XT Delivery System CLIN/TCDS0303-XT, TriClip G4 XTW Delivery System CLIN/TCDS0303-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT. For cardiac procedures.

Codes / lots: Part/UDI-GTIN: CDS0704-NT/08717648234576, CDS0704-NTW/08717648234590, CDS0704-XT/08717648234583, CDS0704-XTW/08717648234606, TCDS0203-NT/08717648229992, TCDS0203-XT/08717648229961, TCDS0303-NT/08717648334276, TCDS0303-NTW/08717648334290, TCDS0303-XT/08717648334283, TCDS0303-XTW/08717648334306, TCDS0203-NT/08717648229992, TCDS0203-XT/08717648229961 All lots.

Reason for recall

An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

FDA-determined cause: Process control

Action taken

On 9/8/22, correction notices were distributed to customers who were asked to do the following: 1) Read and share the information in the notice with all device users. 2) continue to follow the steps of the IFU (eifu.abbottvascular.com) that is also summarized in the correction notice. It details the most relevant steps to aid in optimal function of the Clip locking mechanism, thereby reducing the probability of EFAA failure and COWL to occur. 3) Complete and return the Acknowledgement Form. 4) If you have transferred affected products to other centers, inform those centers of this notice. The recalling firm is working on producing new lots with updated manufacturing processing and raw material to mitigate the EFAA/COWL events. They are also using the correction notice to emphasize the most relevant sections of the IFU that can reduce occurrence. Customers with questions can contact Customer Service at 1800-544-1664.

Details

Recall number
Z-0117-2023
Initiated
September 8, 2022
Posted by FDA
October 20, 2022
Product code
NKM – Mitral Valve Repair Devices
Quantity
767
Distribution
Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK…
Recalling firm
Abbott Vascular, Santa Clara, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0115-2023 Class IIProduct/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip…An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been… 09/08/2022Open, Classified
Z-0116-2023 Class IIProduct/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip NTR Clip…An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been… 09/08/2022Open, Classified