Recall Z-0121-2020

Centurion Medical Products Corporation · initiated August 28, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR

Codes / lots: Batch/Lot: 2019061201

Reason for recall

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

FDA-determined cause: Process control

Action taken

Centurion Medical Products issued Notification via Certified Mail, Return Receipt on 8/28/19 to customers who received affected product. Letter states reason for recall, helath risk and action to take: identify inventory within your possession and cease use immediately. Complete the enclosed accountability record, whether or not you have product, and fax or email at your earliest convenience to the attention of Lisa Carpenter as directed. Upon receipt of your response Centurion Representative will contact you regarding retrieval of the product from your facility, if required. Forward this recall information to any facilities to whom this product may have been further distributed, instructing them to cease use immediately. Questions: Kara A. Kowalczy Director of Quality Assurance(517) 546-5400 Ext.1122. Centurion mailed four (4) additional customer letters on 9/27 /2019 via Certified Mail, Return Receipt.

Details

Recall number
Z-0121-2020
Initiated
August 28, 2019
Terminated
July 23, 2020
Product code
HFX – Clamp, Circumcision
Quantity
204 each
Distribution
Nationwide
Recalling firm
Centurion Medical Products Corporation, Williamston, MI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0120-2020 Class IICenturion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the…A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to… 08/28/2019Terminated