Recall Z-0127-2015

Siemens Healthcare Diagnostics, Inc. · initiated August 14, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dimension Vista HbA1c

Codes / lots: catalog #K3105A, material #10470481, All in-date lots

Reason for recall

In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t

FDA-determined cause: Device Design

Action taken

An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.

Details

Recall number
Z-0127-2015
Initiated
August 14, 2014
Posted by FDA
October 24, 2014
Terminated
August 10, 2015
Product code
LCP – Assay, Glycosylated Hemoglobin
Quantity
20043
Distribution
Worldwide Distribution: US (nationwide) and country of: Canada.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE

Official FDA recall record