Recall Z-0129-2024

B. Braun Medical, Inc. · initiated September 26, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

SA3000 VENTED MICRO SPIKE ADAPTER, REF No. 418121, for intravenous use.

Codes / lots: UDI-DI: 4046964183454 Lots 0061746019 0061755023

Reason for recall

Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous medication. There is also the potential for life-threatening situations to occur requiring additional medical intervention.

FDA-determined cause: Process control

Action taken

On September 26, 2023, the firm notified customers via Urgent Product Recall Notification letters. Customers were informed of the potential for leakage at the male Luer lock site. Customers were instructed to report total amount of impacted product on hand to the firm, after which B Braun Customer Support will contact the customer with instructions on how to return the product for credit or replacement depending on the customer's need. Questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464.

Details

Recall number
Z-0129-2024
Initiated
September 26, 2023
Posted by FDA
November 3, 2023
Product code
FPA – Set, Administration, Intravascular
Quantity
17,100 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Singapore.
Recalling firm
B. Braun Medical, Inc., Allentown, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0128-2024 Class IIFT1000S FLUID TRANSFER SET, REF No. 356048. For direct transfer of fluids between large…Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0130-2024 Class IIAMO ADMINISTRATION SET, Article No. 7A3817, REF OM2505191, for ophthalmologic use.Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0131-2024 Class II30 ML 2PORT BURETTE ASSEMBLY, Article No. 7A3950, used in a contrast management system.Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0132-2024 Class IITERUMO 60" TUBE SET W/HIGH CRACK CHK VLV, Article No. 7S3070. Singapore only. IV…Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0133-2024 Class IITHERASPHERE ADMIN SET, Article No. 7M2650. Canada only. Sterile, single use…Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0134-2024 Class IIALLERGAN UNIVERSAL FILL KIT, REF No. 7M2804. for tissue expander inflation.Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0135-2024 Class II30ML BURETTE ASSY WITH STOPCOCK, BNS, Article No. 7A3970. used in a contrast management…Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified
Z-0136-2024 Class IIBULK LARGEBORE MALE LL FITTING, Article No. M4341750SN. IV administration component.Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase… 09/26/2023Open, Classified