Recall Z-0133-2008

EMD Chemicals Inc · initiated July 30, 2007

Terminated

Product

Harleco May and Grunwald's Stain; 0.17% w/v; Methanol Solution UN1230; EMD Chemicals Inc. An Affilifate of Merck KGaA Damstadt, Germany; 480 South Democrat Road; Gibbstown, NJ 08027; 1 liter and 4 liter poly bottles

Codes / lots: Item # 660/75 Lot 7161(1 liter bottle) Item 660/85, Lot 7161(4 liter bottle) Exp. Date: 31 Jan 09

Reason for recall

Out of Spec -- The Batch (Lot) was incorrectly released after being rejected for out of specification of high azure and eosin absorbance at the end of the run and particulate matter at the end of the run.

FDA-determined cause: Employee error

Action taken

EMD Chemicals via fax and mailed an Important Product Recall letter dated 7/30/07. The letter notified end use consignee that after conducting an investigation it has determined that Harleco May and Grunwald's Stain; 0.17% w/v Item #660/75 % 660/85 has unacceptable levels of azure and eosin absorbance and undissolved dye matter. Users were requested to discontinue shipping and return product. Wholesalers were advised this is a Customer Level Product Recall and requested to sub recall product sold after 7/5/07.

Details

Recall number
Z-0133-2008
Initiated
July 30, 2007
Posted by FDA
November 3, 2007
Terminated
December 26, 2007
Product code
KQC – Stains, Hematology
Quantity
26 units (1 liter), 1 unit (4 liter)
Distribution
Product was released to 2 distributors; one in the US, one in Canada
Recalling firm
EMD Chemicals Inc, Gibbstown, NJ

Official FDA recall record