Recall Z-0137-2024

Maquet Cardiovascular, LLC · initiated September 13, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z

Codes / lots: UDI-DI: 00607567100008 Lot Numbers: 3000188485, 3000193165, 3000193670, 3000195673, 3000198652, 3000199944, 3000201253, 3000204442

Reason for recall

Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.

FDA-determined cause: Device Design

Action taken

Maquet /Getinge issued Urgent Device Medical Device Removal (FSCA 2242352-09/06/2023-004-R) via FedEx with delivery receipt on 9/13/23. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Acrobat-i Stabilizer System with the product codes/lot numbers listed in this notice. If you have unused/unexpired affected product that you will be returning from your inventory, please contact Maquet/Getinge Customer Service at 888-880-2874 between the hours of 6 AM and 5 PM Pacific Standard Time to request a return authorization number (RMA) and shipping instructions. If you have affected product, you are entitled to a credit. You will receive credit upon your acknowledgement that you have affected product for return. Please complete and sign the attached MEDICAL DEVICE Removal - RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by emailing a scanned copy to acrobati-mountbreak2023.act@getinge.com or by faxing the form to (877) 634-6230. Please forward this information to all current and potential Acrobat-i Stabilizer System users within your hospital / facility.

Details

Recall number
Z-0137-2024
Initiated
September 13, 2023
Posted by FDA
October 22, 2023
Product code
MWS – Stabilizer, Heart
Quantity
2951 units
Distribution
Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
Recalling firm
Maquet Cardiovascular, LLC, Wayne, NJ

Official FDA recall record