Recall Z-0138-2011

Instrumentation Laboratory Co. · initiated August 13, 2010

Terminated

Product

ACL TOP CTS PN 000280020 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Codes / lots: Serial Number range: 06020101-10030659

Reason for recall

Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves

FDA-determined cause: Software design

Action taken

Instrumentaion Laboratory notified the US Distributor/Canada via BCI Service Modification Notice (#10793) on August 13, 2010. The Mandatory software upgrade to Verison V4.3.0 in the United States and Canada will be conducted by BCI service representatives and tracked through the distributor, Beckman Coulter (BCI). International Affiliates/Dealers: Worldwide Regulatory Notices were e-mailed to IL international affiliates and dealers, notifying them of the mandatory software upgrade, providing issue description and a risk summary and the support. The upgrades were to be handled through the local representatives.

Details

Recall number
Z-0138-2011
Initiated
August 13, 2010
Posted by FDA
October 25, 2010
Terminated
May 23, 2012
Product code
GKP – Instrument, Coagulation, Automated
Quantity
609 units
Distribution
Worldwide Distribution, USA and Canada
Recalling firm
Instrumentation Laboratory Co., Bedford, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0135-2011 ACL TOP (Base) PN 000280000 Bench top, fully automated, random access analyzer designed…Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot… 08/13/2010Terminated
Z-0136-2011 ACL TOP 500 CTS PN 000280040 Bench top, fully automated, random access analyzer designed…Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot… 08/13/2010Terminated
Z-0137-2011 ACL TOP 700 LAS PN 000280030 Bench top, fully automated, random access analyzer designed…Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot… 08/13/2010Terminated