Recall Z-0138-2023

Merit Medical Systems, Inc. · initiated August 3, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CORCA1410S/CNB, CORCA1415S/B, CORCA1610/CNB, CORCA1615S/CNB, CORCA1810SB/CNB, CORCA1815SB/CNB, CORCA1820SB/CNB, CORCA2015SB/CNB, CORCA2020SB/CNB

Codes / lots: Catalog Code-Lot (OUS Distribution Only): CORA1410/B- I2206909, I2318475, I2291980, I2286704; CORA1410S/B- I2206888; CORA1415/B- I2213361; CORA1615S/B- I2114518, H1517631, H1715532, H1756745, H1767115, H1771762, H1776067, H1784861, H1800979, H1808121, H1825693, I1781878, I1818817, I1837234, I1843704, I1855418, I1866359, I1952354, I1958525, I1971876, I1990075, I1998862, I2043630, I2070249, I2075810; CORA2010S/B- I2010414; CORCA1410S/B- I2259093; CORCA1410S/CNB- I2305358; CORCA1415S/B- I2210950; CORCA1610/CNB- I2291983; CORCA1615S/CNB- I2330683; CORCA1810SB/CNB- I2305359; CORCA1815SB/CNB- I2170254, I2205164; CORCA1820SB/CNB- I2325912; CORCA2015SB/CNB- I2318903, I2325926; CORCA2020SB/CNB- I2320880

Reason for recall

Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.

FDA-determined cause: Process change control

Action taken

On 8/3/2022, recall notices were mailed to customers who were asked to do the following: 1) Identify affected devices, quarantine, discontinue use and distribution, and return to the recalling firm. 2) If product was further distributed, share the recall notice with these customers/users. 3) Complete and return the customer response form. Customers with questions can contact Customer Service at 801-208-4381, 6AM-6PM, Mon-Fri.

Details

Recall number
Z-0138-2023
Initiated
August 3, 2022
Posted by FDA
October 27, 2022
Product code
KNW – Instrument, Biopsy
Quantity
2894
Distribution
Worldwide distribution - US National distribution including in the states of FL, IL, OH, PA, CA, WI, VA, AZ, AR, MS, TN, MO, MA, MI, NY, NE, TX, WA, MN, LA, DE, IN, AL, NJ, OR, MD, GA, CT, OK, UT, KS, NC, CO, KY, WV, AK, NV, RI, SC, ND, NH, MT and the countries of Finland, Denmark, United Arab Emirates, Spain, Sweden, Norway, Jordan, Australia, Germany, Austria, Italy, United Kingdom, France…
Recalling firm
Merit Medical Systems, Inc., South Jordan, UT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0137-2023 Class IICorVocet Biopsy System/Catalog Code: CORA1610/B, CORA1610S/B, CORA1615/B, CORA1810/A…Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly… 08/03/2022Open, Classified