Recall Z-0142-2008

Medtronic Sofamor Danek USA Inc · initiated September 4, 2007

Terminated

Product

Medtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog: 6972267, SIZE 7 X 16, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, Tennessee USA

Codes / lots: Lot Numbers: CN07A034, CN07A031, 434185, 508110, 508111, 508108, 515701, 515702, 508112, 515686, 518409, 518410, 522181, 522179 and 522180

Reason for recall

Implant mis-seating; Variance in size between the trial and the implant could cause the implant to be imporperly seated.

FDA-determined cause: Device Design

Action taken

The firm notified their consignees of the problem and the recall by Urgent Medical Device Recall letter, telephone calls and electronic mail on 09/04/2007. Instructions include requests to immediately quarantine the product and hold for recalling firm collection.

Details

Recall number
Z-0142-2008
Initiated
September 4, 2007
Posted by FDA
December 19, 2007
Terminated
November 3, 2008
Product code
MJO – Prosthesis, Intervertebral Disc
Quantity
196 units
Distribution
Nationwide
Recalling firm
Medtronic Sofamor Danek USA Inc, Memphis, TN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0138-2008 Medtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #:…Implant mis-seating; Variance in size between the trial and the implant could cause the implant to be… 09/04/2007Terminated
Z-0139-2008 Medtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #:…Implant mis-seating; Variance in size between the trial and the implant could cause the implant to be… 09/04/2007Terminated
Z-0140-2008 Medtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #:…Implant mis-seating; Variance in size between the trial and the implant could cause the implant to be… 09/04/2007Terminated
Z-0141-2008 Medtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #:…Implant mis-seating; Variance in size between the trial and the implant could cause the implant to be… 09/04/2007Terminated
Z-0143-2008 Medtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #:…Implant mis-seating; Variance in size between the trial and the implant could cause the implant to be… 09/04/2007Terminated