Recall Z-0146-2010

Hospira Inc · initiated August 11, 2009

Terminated

Product

Power cord for Q2Plus/Continuous Cardiac Output (CCO)/SO2 Computer, list #s 56711-04-01, 56711-04-03, 5671-04-05, 56711-04-51, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.

Codes / lots: AC power cords, all units of Part Numbers: 826-85559-001, 826-85559-003, 826-85559-008, 826-85559-010, 826-96057-003

Reason for recall

Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.

FDA-determined cause: Component design/selection

Action taken

Hospira initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and a letter to consignees was sent via UPS. The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.

Details

Recall number
Z-0146-2010
Initiated
August 11, 2009
Posted by FDA
November 9, 2009
Terminated
April 4, 2012
Product code
DXG – Computer, Diagnostic, Pre-Programmed, Single-Function
Quantity
204,999 power cords distributed for all devices
Distribution
Worldwide distribution.
Recalling firm
Hospira Inc

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0142-2010 Power cord for Model 3300 Cardiac Output Computer, List # 41245-04-03, AC cord…Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock… 08/11/2009Terminated
Z-0143-2010 Power cord for Optional Thermal Printer (For use with Model 3300 COC), Oximetrix 3…Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock… 08/11/2009Terminated
Z-0144-2010 Power cord for Oximetrix 3 SO2/CO Cardiac Output Computer, list # 50130-04-07, AC cord…Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock… 08/11/2009Terminated
Z-0145-2010 Power cord for Q2 Continuous Cardiac Output (CCO)/SO2 Computer, list #s 52235-04-01…Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock… 08/11/2009Terminated
Z-0147-2010 Power cord for QVue Continuous Cardiac Output (CCO)/SO2 computer, list #s 52230-04-03…Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock… 08/11/2009Terminated