Recall Z-0152-2007

Hitachi Medical Systems America Inc · initiated August 14, 2006

Terminated

Product

Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.

Codes / lots: Echelon unit Serial Number: V001. CTL coil component type : MR-CTL-150, N/N KH19636602.

Reason for recall

Component defect/overheating device; A malfunctioning decoupling circuit in the CTL coil caused high current flow and circuit overheating.

FDA-determined cause: Other

Action taken

The recalling firm issued a Device Correction letter to the clinic customer on 9/15/2006 making them aware of this potential problem and of their completed corrective action on 8/14/2006, by a Service Rep visit.

Details

Recall number
Z-0152-2007
Initiated
August 14, 2006
Posted by FDA
November 4, 2006
Terminated
September 19, 2012
Product code
MOS – Coil, Magnetic Resonance, Specialty
Quantity
1 unit
Distribution
The recalled unit was distributed to a single customer.
Recalling firm
Hitachi Medical Systems America Inc, Twinsburg, OH

Official FDA recall record