Recall Z-0158-2015

Terumo Cardiovascular Systems Corp · initiated March 1, 2011

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Terumo Cardioplegia Delivery Set- Item Number: 16130 Product Usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery.

Codes / lots: Lot Numbers: MN08, MNI5, MM11, MM18, MM25. Manufacturing dates October 1I, 2010 through Novemberl5, 20 I 0.

Reason for recall

Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line

FDA-determined cause: Device Design

Action taken

Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.

Details

Recall number
Z-0158-2015
Initiated
March 1, 2011
Posted by FDA
October 31, 2014
Terminated
November 6, 2014
Product code
DTR – Heat-Exchanger, Cardiopulmonary Bypass
Quantity
777 units
Distribution
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Recalling firm
Terumo Cardiovascular Systems Corp, Ashland, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0156-2015 Class IITerumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item…Defect in the luer connector can compromise the seal on the female side of the connector and the compromised… 03/01/2011Terminated
Z-0157-2015 Class IITerumo Custom Integrated Cardioplegia Delivery Set-LINE FOR 15501/16015/5852 Item…Defect in the luer connector can compromise the seal on the female side of the connector. and the… 03/01/2011Terminated
Z-0159-2015 Class IITerumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a…Defect in the luer connector can compromise the seal on the female side of the connector and the compromised… 03/01/2011Terminated
Z-0160-2015 Class IITerumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375…Defect in the luer connector can compromise the seal on the female side of the connector. and the… 03/01/2011Terminated
Z-0161-2015 Class IITerumo Custom Cardioplegia Delivery Sets Catalog Code: 204880 Product Usage: This is a…Defect in the luer connector can compromise the seal on the female side of the connector and the compromised… 03/01/2011Terminated