Recall Z-0164-2007

USA Instruments Incorporated · initiated September 20, 2006

Terminated

Product

3.0T HD Head Neck Spine Array, Model 2415545, for GE 3.0T Excite MR System.

Codes / lots: The following Serial Numbers of this device were subject to recall: 101, 102, 201, 202, 203, 204, 205, 206, 207, 309, 310.

Reason for recall

The firm determined that certain potential conditions for use of their medical device, outside of recommended practices, or operating manual descriptions, could result in a localized RF burn and/or electrical shock to a patient on which the device is being used.

FDA-determined cause: Other

Action taken

The user/customers were notified of this recall with an ''Urgent Customer Safety Notification'' which was sent to the user/customers via Certified mail per recall letter dated 09/20/2006. The notification informs the customer(s) of this potential problem and the three (3) conditions under which this problem (patient becoming burned) could possibly occur. The notification goes on to inform the customers that the firm will be upgrading all 1.5T HD Head Neck & Spine Arrays so as to eliminate this potential hazard, at no cost to the customer and that a Service Representative will be contacting the customer in order to make arrangements to upgrade the coil on-site. A toll-free telelphone no. is provided in the notification letter for questions and/or comments. Lastly, the recall notification provides interim instructions which the customer should follow. The operator/technician should: ask that the patients place their hands at their sides during a scan; immediately discontinue the scan if the patient undergoing the scan starts to move; and stay in constant visual and oral contact with the patient undergoing the scan at all times. For the 3.0T HD Head Neck Spine Array, Model 2415545, for GE 3.0T Excite MR System, The user/customers were notified of this recall with an ''Urgent Customer Safety Notification'' which was sent to the user/customers via Certified mail per recall letter dated 10/13/2006.

Details

Recall number
Z-0164-2007
Initiated
September 20, 2006
Posted by FDA
November 9, 2006
Terminated
April 13, 2012
Product code
MOS – Coil, Magnetic Resonance, Specialty
Quantity
12 distributed in the US.
Distribution
Worldwide, including USA, (states of CA, NJ, PA, MN, IA, MA, WI, WA, and FL) Switzerland, the United Kingdom, France, Canada and Finland.
Recalling firm
USA Instruments Incorporated, Aurora, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0163-2007 1.5T HD Head Neck Spine Array, Model 2416329, for GE 1.5T Excite MR System.The firm determined that certain potential conditions for use of their medical device, outside of… 09/20/2006Terminated