Recall Z-0167-2012

AMO Manufacturing USA, LLC · initiated October 12, 2011

Terminated

Product

Abbott brand Advanced Femtosecond Laser Systems, Model iFS, Catalog Number: J20007D/J/K; Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA Product Usage: Intended Use for: IntraLase Femtosecond (Model 2 and 3) and iFS Advanced Femtosecond Laser Systems. FS Laser Systems are ophthalmic surgical lasers used in certain patient populations: (1) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea; (2) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments; (3) In lamellar IEK [lntraLase Enabled Keratoplasty] and corneal harvesting ; (4) In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea; (5) In the creation of a lamellar cut/resection of the cornea for lamellar IEK for the creation of a penetrating cut/incision for penetrating IEK.

Codes / lots: All units of these models (models 2, 3 and iFS) except model 1.

Reason for recall

The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System.

FDA-determined cause: Process design

Action taken

Abbott Medical Optics Inc. (AMO) issued an "ADVISORY NOTICE" letter dated October 12, 2011 to all affected customers. The letter identifies the affected products, the problem, and the actions needed to be taken. Customers were instructed to exercise caution when making cuts that approach the corneal endothelium. The letter includes a 1-page Operator's Manual addendum that clarifies the recommendations. The letter states that AMO will conduct a field correction to adjust the EOF on the affected FS Laser Systems. The letter includes a response form to be completed and returned as per the instructions. Customers were instructed to notify all appropriate staff/personnel of the Advisory Notice. For any further questions regarding the information or recommendations, contact an AMO representative at 1-800-266-3375

Details

Recall number
Z-0167-2012
Initiated
October 12, 2011
Posted by FDA
November 4, 2011
Terminated
May 1, 2013
Product code
HNO – Keratome, Ac-Powered
Quantity
1226 total laser systems
Distribution
Worldwide Distribution - USA (nationwide)
Recalling firm
AMO Manufacturing USA, LLC, Milpitas, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0165-2012 Abbott brand IntraLase Femtosecond Laser System, Model 2; Catalog Numbers: 20003D/J/K…The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or… 10/12/2011Terminated
Z-0166-2012 Abbott brand IntraLase Femtosecond Laser System, Model 3; Catalog Numbers: 20004D/J/K…The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or… 10/12/2011Terminated