Recall Z-0169-2024

Covidien · initiated August 10, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only

Codes / lots: UDI: 10884521824386/ Lot # 22102001

Reason for recall

Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.

FDA-determined cause: Under Investigation by firm

Action taken

On August 10, 2023 Medtronic ( Parent company of Covidien) issued an Urgent via UPS. Medtronic ask consignees to take the following actions: 1. Quarantine all unused product from the affected lots of McGRATH MAC 2 disposable laryngoscope blades. See attachment A for guidance on identifying potentially affected devices. 2. Return all unused product from the affected lots in your inventory to Medtronic as described on the Customer Confirmation Form. 3. Please complete and return the enclosed Customer Confirmation Form even if you do not have unused inventory. 4. Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed. 5. Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic: . Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or . Call FDA (800) FDA-1088 . Call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1

Details

Recall number
Z-0169-2024
Initiated
August 10, 2023
Posted by FDA
November 1, 2023
Product code
CCW – Laryngoscope, Rigid
Quantity
42,367
Distribution
US: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM NY OH OK OR PA SD TN TX UT VA WA WI WY OUS: Australia Austria Azerbaijan Bahamas Belgium Croatia Czech Republic Denmark Finland France French Guiana Germany Guadeloupe Hong Kong Hungary Ireland Israel Italy Kazakhstan Korea, Republic Of Malta Mayotte Netherlands New Zealand Norway Pakistan Poland Portugal Reunion…
Recalling firm
Covidien, Boulder, CO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0168-2024 Class IIMcGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-017-000, Sterile EO, Rx OnlyLaryngoscope blades may contain an ineffective application of anti-fog agent which may result in… 08/10/2023Open, Classified