Recall Z-0170-2020
GN Otometrics · initiated May 28, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
The ICS CHARTR EP 200
Codes / lots: Part No. Description UDI of the Device 8-04-12733 ICS Chartr EP 200 2Ch, TDH49 , 115/60 05713315006449; 8-04-12734 ICS Chartr EP 200 2Ch,Insert&Bone 05713315006173; 8-04-12731 ICS Chartr EP 200 2ch, 230 VAC (50 Hz) Incl. Insert Earphone; TDH49 Earhone w cable, Bone Conduction Transducer (B71), VEMP Monitor Kit and ASSR 05713315006432; 8-04-12730 ICS Chartr EP 200 2ch, 230 VAC (50 Hz) Incl. Insert Earphone; TDH49 Earhone w cable, Bone Conduction Transducer (B71) and ASSR 05713315006425; 8-04-12729 ICS Chartr EP 200 2Ch, 230 VAC (50 Hz) Incl. Insert Earphone, Bone Conduction Transducer (B71) and VEMP Monitor Kit 05713315006388; 8-04-12727 ICS Chartr EP 200 2Ch, 230 VAC (50 Hz) Incl. Insert Earphone, TDH49 Earhone w cable and VEMP Monitor Kit 05713315006418; 8-04-12725 ICS Chartr EP 200 2ch, 230 VAC (50 Hz) Incl. Insert Earphone, TDH49 Earhone w cable, Bone Conduction Transducer (B71), VEMP Monitor Kit, P300 and ASSR 05713315006395; 8-04-12723 ICS Chartr EP 200 2ch. 230 VAC (50 Hz) Incl. Insert Earphone, TDH49 Earhone w cable 05713315006401; 8-04-12721 ICS Chartr EP 200 2ch. 230 VAC (50 Hz) Incl. Insert Earphone 05713315006371; 8-04-12720 ICS Chartr EP 200 2ch. 230 VAC (50 Hz) Incl.…
Reason for recall
There is a risk to the healthcare professional or patient of exposure to undergrounded electrical surfaces which may result in an electrical shock.
FDA-determined cause: Device Design
Action taken
On May 2019 Otometrics a division of Natus sent letter to the consignees with the following instructions: General Information FSN Type: Recall Natus will be removing all affected products from the market. Further information or advice While there have not been any reported safety incidences, our commitment to providing only the highest quality products and information to our customers and distribution partners is our top priority, and we sincerely apologize for any inconvenience this will cause. If there are any questions about this notice, please contact your authorized Natus distributor. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (As appropriate) Please transfer this notice to other organizations on which this action has an impact. (As appropriate) Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback EU customers only: The Competent (Regulatory) Authority of your country has been informed about this communication to customers. Attached Instructions for device return Customer Reply Form List of affected part numbers INSTRUCTIONS FOR DEVICE RETURN We kindly ask that you initiate the following activities immediately: 1. Identify and quarantine the affected devices. Cease use of all affected devices immediately. 2. Review this communication in its entirety and ensure that all users of the listed products in your organization are informed of this notification. 3. Distribute this notification to all associated parties within your organization. 4. Complete and return the Verification Form enclosed in this letter via e-mail o
Details
- Recall number
- Z-0170-2020
- Initiated
- May 28, 2019
- Terminated
- April 9, 2021
- Product code
- GWJ – Stimulator, Auditory, Evoked Response
- Quantity
- 3071 individual units
- Distribution
- AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cote D'ivoire, Croatia, Czech Republic, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia…
- Recalling firm
- GN Otometrics, Taastrup