Recall Z-0170-2024

Asensus Surgical Inc · initiated September 15, 2023

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system

Codes / lots: UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4

Reason for recall

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.

FDA-determined cause: Software design

Action taken

Asensus Surgical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 09/15/2023 by mail. The notice explained the problem, risk to health, and requested users to cease use of the device until software version 2.7.5 is installed.

Details

Recall number
Z-0170-2024
Initiated
September 15, 2023
Posted by FDA
October 30, 2023
Terminated
September 26, 2024
Product code
NAY – System, Surgical, Computer Controlled Instrument
Quantity
21 units
Distribution
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
Recalling firm
Asensus Surgical Inc, Durham, NC

Official FDA recall record