Recall Z-0175-2007

Asahi Medical Co., Ltd · initiated October 12, 2006

Terminated

Product

Asahi AM-R Series Hollow Fiber Dialyzers; a dry model single or multiple use dialyzer consisting of hollow fiber membranes of cuprammonium rayon housed within a plastic casing of styrene butadiene block polymer, sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Made in Japan; 12 units per case; Model AM-NR-65U - 1.3 m2 surface area, Model AM-NR-75U - 1.5 m2 surface area and Model AM-NR-100U - 2.0 m2 surface area

Codes / lots: Model AM-NR-65U, Model AM-NR-75U and Model AM-NR-100U, all lots

Reason for recall

There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.

FDA-determined cause: Other

Action taken

Asahi sent Safety Alert letters on 10/12/06 to their re-use dialyzer customers advising them of the reports of difficulty separating the blood set from the dialyzer and not to use the Gambro Cartridge Blood Set with Asahi Reuse/Multiple Use Dialyzers in a reuse situation. Enclosed were copies of the temporary revision Instructions For Use with the following warning added at the top of the page: 'Additional Warning for Reuse/Multiple Use Dialyzers: Ashai Reuse/Multiple use dialyzers must not be used in connection with the Gambro cartridge Blood Set in a reuse environment.' Any questions were directed to Karen Borling, RN at 847-498-8500.

Details

Recall number
Z-0175-2007
Initiated
October 12, 2006
Posted by FDA
November 14, 2006
Terminated
November 28, 2007
Product code
FJI – Dialyzer, Capillary, Hollow Fiber
Quantity
100,000 for all models
Distribution
Nationwide
Recalling firm
Asahi Medical Co., Ltd, Northbrook, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0171-2007 Asahi APS Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use dialyzer…There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple… 10/12/2006Terminated
Z-0172-2007 Asahi Rexeed Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use…There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple… 10/12/2006Terminated
Z-0173-2007 Asahi AM-BIO Series Hollow Fiber Dialyzers; a wet model single or multiple use dialyzer…There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple… 10/12/2006Terminated
Z-0174-2007 Asahi AM-BIO D Series Hollow Fiber Dialyzers; a dry model single or multiple use…There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple… 10/12/2006Terminated