Recall Z-0176-2026
American Contract Systems Inc. · initiated September 4, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA75I VASCULAR PACK
Codes / lots: ANCV78BB UDI-DI 191072214362 Lot 8133811 Bag serial number 68273380; ANCV78BC UDI-DI 191072235169 Lot 8499811 Bag serial numbers N/A; HGCV05K UDI-DI 191072212474 Lot 8021411 Bag serial number 68154464; MHPP97AR UDI-DI 191072237446 Lot 8527511 Bag serial number N/A; UDBH37AW UDI-DI 191072235299 Lot 8194311 Bag serial numbers 68198334 68198335; UHPP82AH UDI-DI 191072216434 Lot 8112811 Bag serial number N/A; UICD62AN UDI-DI 191072200495 Lot 7999011 Bag serial number N/A; UICT16F UDI-DI 191072150431 Lot 8000411 Bag serial number N/A Lot 8331311 Bag serial number 17627819 Lot 8429011 Bag serial number 17749736; UIVA75I UDI-DI 191072229052 Lot 8056011 Bag serial number N/A; UIVA75J UDI-DI 191072235602 Lot 8422011 Bag serial number 68667536 * If Bag serial number is "n/a", then all of that lot is affected
Reason for recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
FDA-determined cause: Nonconforming Material/Component
Action taken
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Details
- Recall number
- Z-0176-2026
- Initiated
- September 4, 2025
- Posted by FDA
- October 16, 2025
- Product code
- OEZ – Cardiovascular Procedure Kit
- Quantity
- 32,433 total
- Distribution
- US Nationwide distribution in the states of SD, IA, MN, WA, IL.
- Recalling firm
- American Contract Systems Inc., Grimes, IA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0170-2026 Class II | Basic Biopsy Tray, Item Number NMBP44LIdentified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0171-2026 Class II | Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0172-2026 Class II | Medical convenience kits Item number/ Product description AKGN82E GENERAL AMCB08Y CHEST…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0173-2026 Class II | Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK -…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0174-2026 Class II | Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK -…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0175-2026 Class II | Medical convenience kits Item Number/Description EVLH67F LEFT HEART CATH PACK FDCL22R…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0177-2026 Class II | Medical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0178-2026 Class II | Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0179-2026 Class II | Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0180-2026 Class II | Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0181-2026 Class II | Medical convenience kits Item Number/Description FHCP08Y C SECTION PK HICS50I C-SECTION…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0182-2026 Class II | Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACKIdentified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0183-2026 Class II | Medical convenience kits Item Number/Description HSLM53AB SPINE PACK MILM82A DTC…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0184-2026 Class II | Medical convenience kits Item Number/Description ANSH14AF SHOULDER PACK - 205962 EVAH76H…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0185-2026 Class II | Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI…Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0186-2026 Class II | Medical convenience kits UDPG88W PICC G-TUBE PACK-230209Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |
| Z-0187-2026 Class II | Medical convenience kits IHDC30R D AND C PACKIdentified products were re-gassed (sterilized more than once) after a nonconformance occurred during the… | 09/04/2025Open, Classified |