Recall Z-0185-2019

Alcon Research, LTD. · initiated August 28, 2018

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.

Codes / lots: GTIN: 00380657540037 All lot/serial numbers

Reason for recall

Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.

FDA-determined cause: Device Design

Action taken

The firm initiated their market withdrawal by letter on 08/28/2018. The firm requested surgeons immediately cease implantation of the device and return any unused product.

Details

Recall number
Z-0185-2019
Initiated
August 28, 2018
Posted by FDA
October 23, 2018
Terminated
December 18, 2020
Product code
OGO – Intraocular Pressure Lowering Implant
Quantity
30561 units
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
Recalling firm
Alcon Research, LTD., Sinking Spring, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0186-2019 Class ICyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS…Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent… 08/28/2018Terminated
Z-0187-2019 Class ICyPass ULTRA SYSTEM, CYPASS MICRO-STENT, CYPASS APPLIER, REF CATALOG #: 8065754007. For…Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent… 08/28/2018Terminated
Z-0188-2019 Class ICyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS…Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent… 08/28/2018Terminated