Recall Z-0202-2024

Randox Laboratories Ltd. · initiated September 6, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

RX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000

Codes / lots: GTIN: 05055273207750 Lot/ Batch Number 230T240CS0126S 210T240CS0258S S180T240CS0486MA S180T240CS0487MA S180T240CS0483MA S180T240CS0218MA S170T240CS0665MA S160T240CS0111MA S160T240CS0110MA S150T240CS0005MA S140T240CS0185MA S140T240CS0017MA S140T240CS0015MA

Reason for recall

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

FDA-determined cause: Under Investigation by firm

Action taken

Randox issued Medical Device Correction letter (REC687) on 9/13/23 to to US and PR customers on 9/12/23. Letter states reason for recall, health risk and action to take: Randox recommendations: " Auto rerun settings are disabled, see Appendix A for guidance for each Rx instrument. " Results with flag(<) reported as the bottom of the technical range. Sample Volume settings Low should not be applied. " Results with flag (>) reported as the top of the technical range. Sample Volume settings high and dilutions should not be applied. A manual dilution should be performed, and results multiplied by the dilution factor. The RX Parameter rerun settings will be updated with a parameter release in due course, the Instructions For Use will also be updated at this time. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation.

Details

Recall number
Z-0202-2024
Initiated
September 6, 2023
Posted by FDA
October 31, 2023
Product code
CGA – Glucose Oxidase, Glucose
Quantity
13 units
Distribution
US Nationwide distribution including Puerto Rico
Recalling firm
Randox Laboratories Ltd., Crumlin, North

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0198-2024 Class IIRX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample… 09/06/2023Open, Classified
Z-0199-2024 Class IIRX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R…RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample… 09/06/2023Open, Classified
Z-0200-2024 Class IIRX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample… 09/06/2023Open, Classified
Z-0201-2024 Class IIRX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample… 09/06/2023Open, Classified