Recall Z-0204-2015
Roscoe Medical Inc · initiated October 1, 2014
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Transport Rollator with Padded Seat; Large 8-inch wheels with wheel locks (indoor and outdoor use), fold down footrests, ergonomic handles, curved padded backrest and handy wire basket for storage. Item# 30192 Burgundy each/Item# 30194 Blue each.. Transport Rollator is a wheeled rollator with an additional push bar that can be used to push an end-user for short trips. (Walker, Mechanical)
Codes / lots: 30194, Roscoe Transport Rollator, Blue 30192, Roscoe Transport Rollator, Burgundy MH140401502-MH140401890 MH140500217-MH140500466 MH140406602-MH140406891 MH140407092-MH140407291
Reason for recall
The caster may separate completely from the device frame causing injury to user.
FDA-determined cause: Device Design
Action taken
The firm, Roscoe Medical, issued a "URGENT RECALL NOTICE-Timely Response Required" Quarantine Communication to customers between 09/08/2014 and 09/10/2014. The notice identified affected products as 30194, serial number range MH140407092MH140407291, the problem and actions to be taken. The customers were instructed NOT to use this product and to destroy any affected product that may be in their inventory/control; immediately complete and return the attached Recall Response Form/Certificate of Destruction to Roscoe Medical- "Transport Rollator Recall" within 15 days of receipt of the recall notification via fax to (440) 572-4261 or email: corporate@roscoemedical.com, even if you do not have any affected product; and notify your customers immediately of this product recall. Should you have any questions, contact Roscoe Medical Customer Support at (800) 376-7263 Monday-Friday 8:00 am EST - 5:00 pm EST.
Details
- Recall number
- Z-0204-2015
- Initiated
- October 1, 2014
- Posted by FDA
- November 12, 2014
- Terminated
- February 22, 2017
- Product code
- ITJ – Walker, Mechanical
- Quantity
- 831 units
- Distribution
- Nationwide Distribution: US including states of: AZ, CA, CO, FL, GA, IA, ID, IL, IN, KY, LA< MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI,and WY.
- Recalling firm
- Roscoe Medical Inc, Strongsville, OH