Recall Z-0208-2020

Candela Corporation · initiated September 10, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Canister HFC-134a / 980g, ALUM Product Number: 1600-00-0223 Revision .0C Coolant canister for use with surgical laser.

Codes / lots: Lot # 1904CC1598 1905CC0007 1905CC0178 1905CC0217 1906CC0196 1906CC0172 1907CC0008 1908CC0025 Additional lots added as of 12/6/2019: 1902CC0043 1902CC0143 1811CC0016 1812CC0074 1812CC0133 1901CC0087 1904CC0126 1802CC0092 1802CC0232 1803CC0169 1804CC0073 1805CC0058 1806CC0068 1807CC0009 1807CC0104 1808CC0112 1809CC0070 1810CC0015 1810CC0117 1907CC0187

Reason for recall

The canister neck bushing may become loose and separate from the body of the canister during use. The issue occurs only when the canisters are used in Candela legacy devices that utilize screw-in or threaded interface type DCD modules

FDA-determined cause: Other

Action taken

1. Immediately examine your inventory and quarantine the identified lots of the GentleCool Pro Cryogen Canister. 2. Immediately discontinue use and distribution of the identified lots of the GentleCool Pro Cryogen Canister. 3. Please complete and return the enclosed Acknowledgement Form - GentleCool Pro Canister Recall. The completed form must be scanned and emailed to cryogen.usa@candelamedical.com within 24 hours. 4. If you may have further distributed this product, please identify those locations and notify them at once of this product recall. Your communication should include a copy of this recall notification letter. 5. Please contact Candela Customer Service at 800-733-8550 ext. 0 to make arrangements for replacement of your affected inventory. 6. Information regarding retrieval of your affected inventory will be forthcoming from Candela Medical. If you have any questions, please do not hesitate to call our Customer Service team at 800-733-8550 ext. 0, M-F 8:30 am 8 pm EST, or reach them via email at cryogen.usa@candelamedical.com. An Urgent Medical Device Expanded Recall notification letter dated 12/6/19 was sent to customers. Twenty additional lots have now been identified (listed in bold on the following page), since the initial notification was sent mid-September. Thus, we are updating this notice to make you aware of these additional lots. The reason for this expanded recall is unchanged from the initial communications sent. Actions to Be Taken by the Customer/User: To safely and efficiently implement this recall, we are asking that you please take the following actions: 1. Immediately examine your inventory and quarantine the identified lots of the GentleCool Pro Cryogen Canister. 2. Immediately discontinue use and distribution of the identified lots of the GentleCool Pro Cryogen Canister. 3. Please complete and return the enclosed Acknowledgement Form - GentleCool Pro Canister Recall. The completed form must be sca

Details

Recall number
Z-0208-2020
Initiated
September 10, 2019
Posted by FDA
October 24, 2019
Terminated
October 20, 2020
Product code
GEX – Powered Laser Surgical Instrument
Quantity
in total 7,560 canisters (3240 additional as of 2/5/20)
Distribution
States: NC GA CA MA OH FL IL CO SD NY KS WA MI PA AZ WI TX LA IA MN VA TN SC OR IN WY MD VT ON BC QC NV OUS: Barbados Canada Germany Spain France HR Portugal Ireland Norway Greece NL South Africa Vietnam Hong Kong Japan New Zealand Australia China Korea
Recalling firm
Candela Corporation, Wayland, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0206-2020 Class IICanister HFC-134a /1000g, ALUM Product Number: 1600-00-0218 Revision .0C Coolant…The canister neck bushing may become loose and separate from the body of the canister during use. The issue… 09/10/2019Terminated
Z-0207-2020 Class IICanister HFC-134a /1000g, ALUM, 15 pack Product Number: 1600-00-0219 Revision .0C…The canister neck bushing may become loose and separate from the body of the canister during use. The issue… 09/10/2019Terminated
Z-0209-2020 Class IICanister HFC-134a /980g, ALUM, 15 pack Product Number: 1600-00-0224 Revision .0C Coolant…The canister neck bushing may become loose and separate from the body of the canister during use. The issue… 09/10/2019Terminated