Recall Z-0212-2024

Baxter Healthcare Corporation · initiated October 4, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974

Codes / lots: UDI/DI 00887761966925, all serial numbers

Reason for recall

Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.

FDA-determined cause: Under Investigation by firm

Action taken

Baxter issued an Urgent Medical Device Correction notice to its consignees on 10/04/2023 by USPS first class mail. The notice explained the potential problem and hazard. Baxter will be correcting this issue by installing a conductive drag chain on the Progressa+ bed frame to prevent the electrostatic charge buildup. A Baxter representative will contact your facility to schedule a device correction to add the conductive drag chain. Please note you will be receiving this correction from Baxter at no charge. All units purchased in the future will have the drag chain.

Details

Recall number
Z-0212-2024
Initiated
October 4, 2023
Posted by FDA
November 2, 2023
Product code
IOQ – Bed, Flotation Therapy, Powered
Quantity
499 beds
Distribution
US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA
Recalling firm
Baxter Healthcare Corporation, Deerfield, IL

Official FDA recall record