Recall Z-0228-2026

Synthes (USA) Products LLC · initiated September 26, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.

Codes / lots: Product Code: 04.045.346TS. GTIN: 10886982301413. Lot Number: 31756P7. Expiration Date: 08/31/2025

Reason for recall

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

FDA-determined cause: Process control

Action taken

DePuy Synthes notified consignees via email and mailed letter on 09/26/2025. Consignees were instructed to examine inventory and quarantine affected units, arrange for the return of any affected units, complete and return the Business Response Form, and place the notification in a visible area for awareness. If units were further distributed within your facility or to customers, please notify those customers and personnel.

Details

Recall number
Z-0228-2026
Initiated
September 26, 2025
Posted by FDA
October 21, 2025
Product code
JDS – Nail, Fixation, Bone
Quantity
120 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
Recalling firm
Synthes (USA) Products LLC, West Chester, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0229-2026 Class IILOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code:…Although the etched length on the screw is correct, the length marked on the external packaging is… 09/26/2025Open, Classified