Recall Z-0236-2018
Beckman Coulter Inc. · initiated November 3, 2017
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
UniCel DxH800, Part Numbers: 629029 and B24802, automated hematology analyzer
Codes / lots: Serial Numbers: B24802: AW10161, AW10154, AZ34449, AZ34447 629029: AT40563, AT06096, AT06095, AT02044, AW05080, AW05072, AW05066, AT50679, AS49422, AS36312, AS23151
Reason for recall
The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
FDA-determined cause: Device Design
Action taken
The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."
Details
- Recall number
- Z-0236-2018
- Initiated
- November 3, 2017
- Terminated
- June 19, 2018
- Product code
- GKZ – Counter, Differential Cell
- Quantity
- 15 units
- Distribution
- CA, AR, HI, CA, OH, WA, MD, Canada, Japan
- Recalling firm
- Beckman Coulter Inc., Brea, CA