Recall Z-0246-2020

Alto Development Corp · initiated September 10, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

A&E Medical Thorecon Plating System Kits Model No. 94-1700-10. For fracture fixation or reconstruction of the sternum.

Codes / lots: 327150, 327151, 327152, 327707, 340284, 343232, 343236

Reason for recall

Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.

FDA-determined cause: Process control

Action taken

On September 10, 2019, the firm distributed a letter to customers titled Product Recall: Action Required. The letter informed customers of the potential tensioner failure. Customers were given two options: 1. A&E will provide a supply of individually packaged, sterile tensioners to be kept on hand should a kit tensioner be found inoperable. Surgeons and Operating Room staff need to be alerted to this situation and the stock of additional tensioners readily available to them during surgery. 2. For those customers who prefer to return their affected lots of Thorecon, A&E will provide a Returned Goods Authorization and replacement product (to be provided as soon as inventory is available). Any failed tensioners can be discarded. Customers were asked to complete and return the acknowledgement form included with the mailing. Should you have any questions about the action or the product, please contact your A&E Sales Representative or our Customer Service at 800-323-4035.

Details

Recall number
Z-0246-2020
Initiated
September 10, 2019
Terminated
May 12, 2020
Product code
JDQ – Cerclage, Fixation
Quantity
209 total
Distribution
Distributed to AL, CA, FL, GA, IL, LA, MO, OK, MS, NM, OK, SC, TX, UT, VT, WA, and WI.
Recalling firm
Alto Development Corp, Wall Township, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0242-2020 Class IIA&E Medical Thorecon Plating System Kits Model No. 94-1200-08. For fracture fixation or…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… 09/10/2019Terminated
Z-0243-2020 Class IIA&E Medical Thorecon Plating System Kits Model No. 94-1300-04-S. For fracture fixation…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… 09/10/2019Terminated
Z-0244-2020 Class IIA&E Medical Thorecon Plating System Kits Model No. 94-1400-04. For fracture fixation or…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… 09/10/2019Terminated
Z-0245-2020 Class IIA&E Medical Thorecon Plating System Kits Model No. 94-1500-04-S. For fracture fixation…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… 09/10/2019Terminated