Recall Z-0246-2020
Alto Development Corp · initiated September 10, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
A&E Medical Thorecon Plating System Kits Model No. 94-1700-10. For fracture fixation or reconstruction of the sternum.
Codes / lots: 327150, 327151, 327152, 327707, 340284, 343232, 343236
Reason for recall
Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.
FDA-determined cause: Process control
Action taken
On September 10, 2019, the firm distributed a letter to customers titled Product Recall: Action Required. The letter informed customers of the potential tensioner failure. Customers were given two options: 1. A&E will provide a supply of individually packaged, sterile tensioners to be kept on hand should a kit tensioner be found inoperable. Surgeons and Operating Room staff need to be alerted to this situation and the stock of additional tensioners readily available to them during surgery. 2. For those customers who prefer to return their affected lots of Thorecon, A&E will provide a Returned Goods Authorization and replacement product (to be provided as soon as inventory is available). Any failed tensioners can be discarded. Customers were asked to complete and return the acknowledgement form included with the mailing. Should you have any questions about the action or the product, please contact your A&E Sales Representative or our Customer Service at 800-323-4035.
Details
- Recall number
- Z-0246-2020
- Initiated
- September 10, 2019
- Terminated
- May 12, 2020
- Product code
- JDQ – Cerclage, Fixation
- Quantity
- 209 total
- Distribution
- Distributed to AL, CA, FL, GA, IL, LA, MO, OK, MS, NM, OK, SC, TX, UT, VT, WA, and WI.
- Recalling firm
- Alto Development Corp, Wall Township, NJ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0242-2020 Class II | A&E Medical Thorecon Plating System Kits Model No. 94-1200-08. For fracture fixation or…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… | 09/10/2019Terminated |
| Z-0243-2020 Class II | A&E Medical Thorecon Plating System Kits Model No. 94-1300-04-S. For fracture fixation…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… | 09/10/2019Terminated |
| Z-0244-2020 Class II | A&E Medical Thorecon Plating System Kits Model No. 94-1400-04. For fracture fixation or…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… | 09/10/2019Terminated |
| Z-0245-2020 Class II | A&E Medical Thorecon Plating System Kits Model No. 94-1500-04-S. For fracture fixation…Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which… | 09/10/2019Terminated |