Recall Z-0270-2012

Uplift Technologies Inc. · initiated June 27, 2011

Terminated

Product

"***UPLIFT TECHNOLOGIES INC.***Bringing quality to life.***Model No.***CA200***Description***Lifting commode chair***Weight Range***80-300+ lb-136+ kg***" Labeling on product: "***1C0101 and K0061014***". A lifting commode chair that is self-powered (non-electric) and provides lifting assistance for those who have reduced mobility due to arthritis, post-operative conditions, degenerative joint disease, Parkinson's disease or other neuromuscular disorders.

Codes / lots: Model CA200 Lot 1C0101.

Reason for recall

On June 26, 2011 Uplift Technologies Inc. Nova Scotia, Canada initiated a recall of the Uplift Commode Assist Model No. CA200, Lot No. 1C0101. The plastic seat does not meet manufacturing specifications. The seat may not fit the unit properly and may interfere with the unit's normal locking mechanism.

FDA-determined cause: Nonconforming Material/Component

Action taken

Uplift Technologies, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 27, 2011 to all affected customers The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to examine their inventory and to quarantine the affected product. Customers are to call Uplift's Customer Service Department at 1-800-387-0896 to request a Return Materials Authorization number in order to return the product. .

Details

Recall number
Z-0270-2012
Initiated
June 27, 2011
Posted by FDA
November 22, 2011
Terminated
May 31, 2012
Product code
INN – Chair, Adjustable, Mechanical
Quantity
219 chairs
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Recalling firm
Uplift Technologies Inc., Dartmouth

Official FDA recall record