Recall Z-0278-2018

Philips Electronics North America Corporation · initiated September 28, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Automated External Defibrillator - HeartStart MRx Monitor/Defibrillator

Codes / lots: 20 units - US00597060, US00597964, US00598048, US00598210, US00598784, US00599490, US00599495, US00599508, US00599572, US00599613, US00599615, US00599623, US00599726, US00600362, US00600363, US00600370, US00600422, US00600423, US00600438, US00600483

Reason for recall

Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).

FDA-determined cause: Nonconforming Material/Component

Action taken

The Philips Urgent Medical Device Correction notification/ Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of providing a hardware correction, free of charge, for all units affected by this issue.

Details

Recall number
Z-0278-2018
Initiated
September 28, 2017
Terminated
October 12, 2018
Product code
MSX – System, Network And Communication, Physiological Monitors
Quantity
20 units ( 18 domestic, 2 foreign)
Distribution
20 units affected. 18 domestic, 2 foreign.
Recalling firm
Philips Electronics North America Corporation, Andover, MA

Official FDA recall record