Recall Z-0278-2026

Sedecal SA · initiated June 10, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Model Number SM-40HF-B-D-C; 40KW.002, Mobile X-ray system

Codes / lots: UDI/DI 08436046001510, Serial Numbers: G39544, G62558, G71736, G40116, G65568, G65570.

Reason for recall

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

FDA-determined cause: Device Design

Action taken

Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.

Details

Recall number
Z-0278-2026
Initiated
June 10, 2025
Posted by FDA
October 23, 2025
Product code
IZL – System, X-Ray, Mobile
Quantity
6 units
Distribution
US: CA, IL, and NJ
Recalling firm
Sedecal SA, Algete

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0242-2026 Class IISedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0243-2026 Class IIMobileDiagnost wDR 2.2; Model Number REF 9890-010-89522, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0244-2026 Class IIModel Number 40KWFX, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0245-2026 Class IIModel Number 40KW-FX.899, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0246-2026 Class IIModel Number 40KWFXPLUS, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0247-2026 Class IIModel Number 40KWFXPLUS.002, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0248-2026 Class IIModel Number 40KWFXPLUS.004, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0249-2026 Class IIModel Number 40KWFXPLUS.005, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0250-2026 Class IIModel Number 40KWFXPLUS.889, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0251-2026 Class IIModel Number 40KWFXPLUS-710CW, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0252-2026 Class IIModel Number 40KWFXPLUS-710CW.007, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0253-2026 Class IIModel Number 40KWMOBILE3.899, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0254-2026 Class IIModel Number SM-20HF-Batt; 20KW ANALOG, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0255-2026 Class IIModel Number SM-32HF-B-D-C; 32KW 50G, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0256-2026 Class IIModel SM-32HF-Batt; 32KW ANALOG, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0257-2026 Class IIModel SM-40HF-B-D-C; 40KW - CANON READY, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0258-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 50G, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0259-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0260-2026 Class IIModel SM-40HF-B-D-C; 40KW 70C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0261-2026 Class IIModel Number SM-40HF-Batt; 40KW ANALOG, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0262-2026 Class IISedecal Easy Moving Plus Model Number SM-50HF-B-D-C; 50KW 50G, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0263-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 50C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0264-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 50G, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0265-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0266-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0267-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55G, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0268-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55G/60G, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0269-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0270-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified
Z-0271-2026 Class IIModel Number SM-40HF-B-D-C; 40KW 70C, Mobile X-ray systemSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and… 06/10/2025Open, Classified