Recall Z-0296-2017

Ethicon Endo-Surgery Inc · initiated July 15, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses

Codes / lots: N91L2N, N91L2P, N91N1J, N91N1K, N91L0T, N91L0U, N91N29, N91N2G, N91R8X, N91R8Y.

Reason for recall

Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.

FDA-determined cause: Device Design

Action taken

Ethicon sent an Urgent Medical Device Recall letter dated July 15, 2016, to all affected consignees via UPS next day mail. The letter identified the product, the problem, and the action to be taken by the customer. Consignees were instructed to examine their inventory immediately and quarantine any affected product and return to ENDOPATH. If product subject to the recall was further distributed consignees were instructed to notify the other facility and arrange for return of the affected product. Customer were instructed to complete the Business Reply Form confirming receipt of this letter and fax to Stericycle at 1-888-228-5719 or email to ethicon5182@stericycle.com. Customers with questions were instructed to contact the Ethicon Customer Support Center at 1-877-384-4266. For questions regarding this recall call 513-337-3582.

Details

Recall number
Z-0296-2017
Initiated
July 15, 2016
Posted by FDA
October 20, 2016
Terminated
February 28, 2017
Product code
GDW – Staple, Implantable
Quantity
213
Distribution
Worldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.
Recalling firm
Ethicon Endo-Surgery Inc, Blue Ash, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0295-2017 Class IIEndoscopic Articulating Linear Cutter, ATB35. Standard 35 MM. The ENDOPATH ETS-Flex…Quality control identified a component issue where the pinion gear in the device could fail under extreme… 07/15/2016Terminated